Multiple detector-row CT angiography of the renal and mesenteric vessels

被引:60
作者
Fleischmann, D [1 ]
机构
[1] Univ Vienna, AKH, Dept Angiog & Intervent Radiol, A-1090 Vienna, Austria
关键词
computed tomographyangiography; intestinesischemia; mesenterycomputed tomography; renal arteriescomputed tomography; stenosis or obstruction;
D O I
10.1016/S0720-048X(02)00364-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computed tomography angiography (CTA) of the abdomen with multiple detector-row computed tomography (MD-CT) is an effective technique for minimally invasive imaging of the renal arteries and the visceral vasculature. This article reviews the clinical and technical aspects of MD-CT angiography in terms of image acquisition and reconstruction parameters, contrast medium application, and three-dimensional visualization with special attention to renal and mesenteric vascular imaging. Because of its high sensitivity to detect renal artery stenosis on the one hand, and because a normal renal CTA virtually excludes the presence of a significant renal artery stenosis on the other hand, renal CTA plays a useful role in the management of patients with suspected renovascular hypertension. Mesenteric CTA is a useful tool for visualizing normal vascular anatomy and its variants-particularly in the setting of organ transplantation. Vascular pathology, e.g. atherosclerotic disease (abdominal angina), or aneurysms of the visceral arteries are reliably assessed with CTA. Mesenteric CTA is an invaluable adjunct to abdominal CT in the setting of abdominal emergencies, because of its ability to detect the causes of acute intestinal ischemia (Superior mesenteric artery embolism or thrombosis, superior mesenteric vein thrombosis). Accurate timing of the CTA acquisition and the subsequent parenchymal phase acquisition relative to the contrast medium transit time is critical to obtain excellent image quality in double-pass abdominal CT acquisitions. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:S79 / S87
页数:9
相关论文
共 16 条
[1]   Helical CT angiography compared with arteriography in the detection of renal artery stenosis [J].
Beregi, JP ;
Elkohen, M ;
Deklunder, G ;
Artaud, D ;
Coullet, JM ;
Wattinne, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) :495-501
[2]   MESENTERIC ISCHEMIA - A MULTIDISCIPLINARY APPROACH [J].
BRADBURY, AW ;
BRITTENDEN, J ;
MCBRIDE, K ;
RUCKLEY, CV .
BRITISH JOURNAL OF SURGERY, 1995, 82 (11) :1446-1459
[3]   Mesenteric venous thrombosis: Diagnosis and noninvasive imaging [J].
Bradbury, MS ;
Kavanagh, PV ;
Bechtold, RE ;
Chen, MY ;
Ott, DJ ;
Regan, JD ;
Weber, TM .
RADIOGRAPHICS, 2002, 22 (03) :527-541
[4]   Use of high concentration contrast media: principles and rationale - vascular district [J].
Fleischmann, D .
EUROPEAN JOURNAL OF RADIOLOGY, 2003, 45 :S88-S93
[5]   Multiphase hepatic CT with a multirow detector CT scanner [J].
Foley, WD ;
Mallisee, TA ;
Hohenwalter, MD ;
Wilson, CR ;
Quiroz, FA ;
Taylor, AJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (03) :679-685
[6]   Special focus session - Multidetector CT: Abdominal visceral imaging [J].
Foley, WD .
RADIOGRAPHICS, 2002, 22 (03) :701-719
[7]   Three-dimensional helical CT angiography in renal transplant recipients: A new problem-solving tool [J].
Hofmann, LV ;
Smith, PA ;
Kuszyk, BS ;
Kraus, E ;
Fishman, EK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (04) :1085-1089
[8]   Multi-detector row CT of mesenteric ischemia: Can it be done? [J].
Horton, KM ;
Fishman, EK .
RADIOGRAPHICS, 2001, 21 (06) :1463-1473
[9]   Renal artery stenosis: CT angiography - Comparison of real-time volume-rendering and maximum intensity projection algorithms [J].
Johnson, PT ;
Halpern, EJ ;
Kuszyk, BS ;
Heath, DG ;
Wechsler, RJ ;
Nazarian, LN ;
Gardiner, GA ;
Levin, DC ;
Fishman, EK .
RADIOLOGY, 1999, 211 (02) :337-343
[10]   Renal artery stenosis: Detection and quantification with spiral CT angiography versus optimized digital subtraction angiography [J].
Kaatee, R ;
Beek, FJA ;
deLange, EE ;
vanLeeuwen, MS ;
Smits, HFM ;
vanderVen, PJG ;
Beutler, JJ ;
Mali, WPTM .
RADIOLOGY, 1997, 205 (01) :121-127