Flow-injection analysis-fluorescence detection for the in vivo on-line determination of calcium in blood with microdialysis sampling

被引:20
作者
Huang, YM
Zhang, ZJ [1 ]
Lv, JG
Cheng, H
机构
[1] SW Normal Univ, Dept Chem, Inst Analyt Sci, Chongqing 400715, Peoples R China
[2] Nanjing Univ, Dept Chem, Nanjing 210008, Peoples R China
基金
中国国家自然科学基金;
关键词
calcium; quantitative microdialysis; in vivo analysis; fluorescence;
D O I
10.1016/S0003-2670(00)01009-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Due to its high selectivity for calcium, calcein can be used as fluorescent indicator for the sensitive determination of calcium. A flow-injection on-line microdialysis system for in vivo monitoring of calcium in blood is developed. A microdialysis probe, implanted in the vein of rabbit, was perfused with perfusate at a flow rate of 5 mu l/min. The concentration of calcium in the dialysate was determined on-line with a flow-injection fluorescence system. The dialysate sample volume was about 15 mu l. The system was linearly related to the concentration of calcium in the range 0-80 mu g/ml (r(2)=0.9968) with a detection limit (3 sigma) of 0.01 mu g/ml and the sample throughput of 20 h(-1). R.S.D. was <5% (n=11). The validation of using the no net flux method of quantitative microdialysis to measure concentration of calcium was evaluated in vitro and in vivo. The accuracy of this method was tested in vitro under steady-state conditions for three test solutions containing known concentration of calcium. The estimated concentration of calcium were 32+/-2, 42+/-2, and 49.5+/-0.9 mu g/ml (mean+/-R.S.D.) in agreement with the actual concentrations of calcium in the test solutions which were 30, 40 and 50 mu g/ml, respectively. In vivo, the estimated basal of calcium in dialysis sample from probe in the test animal's blood was 33 mu g/ml. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:175 / 184
页数:10
相关论文
共 27 条
[1]   MICRODIALYSIS SAMPLING COUPLED ONLINE TO FAST MICROBORE LIQUID-CHROMATOGRAPHY [J].
CHEN, AQ ;
LUNTE, CE .
JOURNAL OF CHROMATOGRAPHY A, 1995, 691 (1-2) :29-35
[2]   A review of microdialysis sampling for pharmacokinetic applications [J].
Davies, MI .
ANALYTICA CHIMICA ACTA, 1999, 379 (03) :227-249
[3]   Simultaneous microdialysis sampling from multiple sites in the liver for the study of phenol metabolism [J].
Davies, MI ;
Lunte, CE .
LIFE SCIENCES, 1996, 59 (12) :1001-1013
[4]  
DETERDING LJ, 1992, ANAL CHEM, V64, P1636
[5]  
Devineni D, 1996, CANCER RES, V56, P1983
[6]   ESTIMATION OF UNBOUND CONCENTRATIONS OF MORPHINE FROM MICRODIALYSATE CONCENTRATIONS BY USE OF NONLINEAR-REGRESSION ANALYSIS INVIVO AND INVITRO DURING STEADY-STATE CONDITIONS [J].
EKBLOM, M ;
GARDMARK, M ;
HAMMARLUNDUDENAES, M .
LIFE SCIENCES, 1992, 51 (06) :449-460
[7]   A flow injection microdialysis sampling chemiluminescence system for in vivo on-line monitoring of glucose in intravenous and subcutaneous tissue fluid microdialysates [J].
Fang, Q ;
Shi, XT ;
Sun, YQ ;
Fang, ZL .
ANALYTICAL CHEMISTRY, 1997, 69 (17) :3570-3577
[8]  
GORTH L, 1996, ACTA DERM-VENER S, V197, P1
[9]  
GRYNKIEWICZ G, 1985, J BIOL CHEM, V260, P3440
[10]   LIQUID-CHROMATOGRAPHIC DETERMINATION OF TACRINE AND ITS METABOLITES IN RAT BILE MICRODIALYSATES [J].
HADWIGER, ME ;
TELTINGDIAZ, M ;
LUNTE, CE .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 655 (02) :235-241