Detection of body position changes using the surface electrocardiogram

被引:16
作者
Åström, M
García, J
Laguna, P
Pahlm, O
Sörnmo, L [1 ]
机构
[1] Lund Univ, Dept Electrosci, Signal Proc Grp, Lund, Sweden
[2] Univ Zaragoza, Aragon Inst Engn Res IZA, Grp Tecnol Comunicac, Zaragoza, Spain
[3] Lund Univ, Dept Clin Physiol, Lund, Sweden
关键词
body position; ischaemia monitoring; loop alignment; rotation angles; VCG loops; VECTORCARDIOGRAPHIC LOOP ALIGNMENT; TRANSIENT MYOCARDIAL-ISCHEMIA; TO-BEAT VARIABILITY; SEGMENT; ECG; ACCELEROMETRY; VALIDATION;
D O I
10.1007/BF02344884
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A method for detecting body position changes that uses the surface vectorcardiogram (VCG) is presented. Such changes are often manifested as sudden shifts in the electrical axis of the heart and can erroneously be interpreted as acute ischaemic events. Axis shifts were detected by analysing the rotation angles obtained from the alignment of successive VCG loops to a reference loop. Following the rejection of angles originating from noise events, the detection of body position changes was performed on the angle series using a Bayesian approach. On a database of ECG recordings from normal subjects performing a predefined sequence of body position changes, a detection rate of 92% and a false alarm rate of 7% was achieved.