Method to predict and compare the influence of the particle size on the isocratic peak capacity of high-performance liquid chromatography columns

被引:34
作者
Cabooter, Deirdre
de Villiers, Andre
Clicq, David
Szucs, Roman
Sandra, Pat
Desmet, Gert
机构
[1] Vrije Univ Brussel, Dept Chem Engn, CHIS TW, B-1050 Brussels, Belgium
[2] Univ Ghent, Pfizer Analyt Res Ctr, B-9000 Ghent, Belgium
[3] Pfizer Global R&D, Analyt R&D, Sandwich, Kent, England
关键词
kinetic plot; isocratic peak capacity; particle size;
D O I
10.1016/j.chroma.2007.02.055
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A kinetic plot based method has been used to experimentally predict the optimal particle size yielding the maximal isocratic peak capacity in a given analysis time. Applying the method to columns of three different manufacturers and characterizing them by separating a 4-component paraben mixture at 30 degrees C, it was consistently found that the classical 3 and 3.5 mu m particles provide the highest peak capacity for typical isocratic separation run times between 30 and 60 min when operating the columns at a conventional pressure of 400 bar. Even at 1000 bar, the sub-2 mu m particles only have a distinct advantage for runs lasting 30 min or less, while for runs lasting 45 min or longer the 3 and 3.5 mu m again are to be preferred. This finding points at the advantage for high-resolution separations that could be obtained by producing 3 and 3.5 mu m particle columns that can be operated at elevated pressures. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:183 / 191
页数:9
相关论文
共 17 条
[1]   FACTORS DETERMINING FLOW-RATE IN CHROMATOGRAPHIC COLUMNS [J].
CRAMERS, CA ;
RIJKS, JA ;
SCHUTJES, CPM .
CHROMATOGRAPHIA, 1981, 14 (07) :439-444
[2]   Evaluation of ultra performance liquid chromatography -: Part I.: Possibilities and limitations [J].
de Villiers, Andre ;
Lestremau, Francois ;
Szucs, Roman ;
Gelebart, Sylvie ;
David, Frank ;
Sandra, Pat .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1127 (1-2) :60-69
[3]   Geometry-independent plate height representation methods for the direct comparison of the kinetic performance of LC supports with a different size or morphology [J].
Desmet, G ;
Clicq, D ;
Gzil, P .
ANALYTICAL CHEMISTRY, 2005, 77 (13) :4058-4070
[4]   Effect of temperature in reversed phase liquid chromatography [J].
Guillarme, D ;
Heinisch, S ;
Rocca, JL .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1052 (1-2) :39-51
[5]   ULTIMATE LIMITS IN HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
HALASZ, I ;
ENDELE, R ;
ASSHAUER, J .
JOURNAL OF CHROMATOGRAPHY, 1975, 112 (OCT29) :37-60
[6]  
Jerkovich AD, 2005, LC GC N AM, P15
[7]   High-efficiency liquid chromatography on conventional columns and instrumentation by using temperature as a variable I.: Experiments with 25 cm x 4.6 mm ID, 5 μm ODS columns [J].
Lestremau, F ;
Cooper, A ;
Szucs, R ;
David, F ;
Sandra, P .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1109 (02) :191-196
[8]   Accuracy of empirical correlations for estimating diffusion coefficients in aqueous organic mixtures [J].
Li, JW ;
Carr, PW .
ANALYTICAL CHEMISTRY, 1997, 69 (13) :2530-2536
[9]  
Majors RE, 2005, LC GC N AM, V23, P1248
[10]   Effect of high-temperature on high-performance liquid chromatography column stability and performance under temperature-programmed conditions [J].
Marin, SJ ;
Jones, BA ;
Felix, WD ;
Clark, J .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1030 (1-2) :255-262