Reproducibility of serial intravascular ultrasound measurements in patients with angiographically silent coronary artery disease after heart transplantation

被引:22
作者
Bocksch, W
Wellnhofer, E
Schartl, M
Dreysse, S
Klimek, W
Franke, R
Musci, M
Hetzer, R
Fleck, E
机构
[1] Humboldt Univ, Dept Internal Med Cardiol, Div Internal Med Cardiol, Cardiac Catheterizat Lab, Berlin, Germany
[2] Humboldt Univ, German Heart Inst Berlin, Berlin, Germany
[3] German Heart Inst Berlin, Dept Cardiothorac & Vasc Surg, Berlin, Germany
关键词
intracoronary ultrasound; reproducibility; transplant vasculopathy;
D O I
10.1097/00019501-200010000-00007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Intracoronary ultrasound (ICUS) imaging is the most sensitive method for the early detection and serial evaluation of vasculopathy of transplants. Both lack of agreement between observers and lack of agreement between serial, independent pullback procedures (repeatability), which can result in a variable intraluminal catheter position may limit the reproducibility of ICUS measurements. Objective To evaluate the reproducibility of serial measurements of standard linear and area cross-sectional coronary dimensions in patients with non-obstructive transplant vasculopathy. Methods We performed ICUS imaging of patients without angiographic evidence of obstructive epicardial coronary artery disease after heart transplantation. A 30 MHz phased-array transducer was used. Two independent pullbacks of the left anterior descending coronary artery were performed and recorded on CD-ROM for off-line quantitative analysis of the most severely diseased site. Agreement of observers and repeatability of serial measurements were calculated by the use of linear regression analysis and Bland-Altman plots. Results Regarding agreement of observers, correlation coefficients for intra-observer agreement ranged from r = 0.98 to r = 0.99; those for interobserver agreement ranged from r = 0.87 to r = 0.98. Serial measurements of the identical coronary artery cross-section within independent catheter pullback procedures were possible for 104 of 112 target lesions (92.9%). Correlation coefficients ranged from r = 0.91 to r = 0.97 (for lumen diameter r = 0.91, for lumen area r = 0.93, for vessel diameter r = 0.91, for vessel area r = 0.97, for thickness of plaque r = 0.96 and for area of plaque 0.94). The mean difference of measurements was around zero for all parameters with SD from 0.13 to 0.4 mm for linear parameters and from 1.53 to 1.82 mm(2) for area parameters. Conclusion Serial intravascular ultrasound measurements are highly reproducible without any evidence of systematic error and a SD of differences of measurements beyond the maximal spatial resolution of currently available intravascular ultrasound catheters. Coron Artery Dis 11:555-562 (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:555 / 562
页数:8
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Safety of intracoronary ultrasound imaging in patients with acute myocardial infarction [J].
Bocksch, W ;
Schartl, M ;
Beckmann, S ;
Dreysse, S ;
Fleck, E .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 81 (05) :641-643
[3]  
Costanzo MR, 1996, J HEART LUNG TRANSPL, V15, P221
[4]   Clinical application and image interpretation in intracoronary ultrasound [J].
Di Mario, C ;
Gorge, G ;
Peters, R ;
Kearney, P ;
Pinto, F ;
Hausmann, D ;
von Birgelen, C ;
Colombo, A ;
Mudra, H ;
Roelandt, J ;
Erbel, R .
EUROPEAN HEART JOURNAL, 1998, 19 (02) :207-229
[5]  
ERBEL R, 1998, INTRAVASCULAR ULTRAS, P23
[6]  
GE J, 1991, Z KARDIOL, V80, P595
[7]  
Geselschap JH, 1998, J ENDOVASC SURG, V5, P126, DOI 10.1583/1074-6218(1998)005<0126:TEOAOI>2.0.CO
[8]  
2
[9]   INTRACORONARY ULTRASOUND IMAGING - INTRAOBSERVER AND INTEROBSERVER VARIABILITY OF MORPHOMETRIC MEASUREMENTS [J].
HAUSMANN, D ;
LUNDKVIST, AJS ;
FRIEDRICH, GJ ;
MULLEN, WL ;
FITZGERALD, PJ ;
YOCK, PG .
AMERICAN HEART JOURNAL, 1994, 128 (04) :674-680
[10]   THE SAFETY OF INTRACORONARY ULTRASOUND - A MULTICENTER SURVEY OF 2207 EXAMINATIONS [J].
HAUSMANN, D ;
ERBEL, R ;
ALIBELLICHEMARIN, MJ ;
BOKSCH, W ;
CARACCIOLO, E ;
COHN, JM ;
CULP, SC ;
DANIEL, WG ;
DESCHEERDER, I ;
DIMARIO, C ;
FERGUSON, JJ ;
FITZGERALD, PJ ;
FRIEDRICH, G ;
GE, JB ;
GORGE, G ;
HANRATH, P ;
HODGSON, JM ;
ISNER, JM ;
JAIN, S ;
MAIERRUDOLPH, W ;
MOONEY, M ;
MOSES, JW ;
MUDRA, H ;
PINTO, FJ ;
SMALLING, RW ;
TALLEY, JD ;
TOBIS, JM ;
WALTER, PD ;
WEIDINGER, F ;
WERNER, GS ;
YEUNG, AC ;
YOCK, PG .
CIRCULATION, 1995, 91 (03) :623-630