On-line preconcentration of lead and cadmium for flame atomic absorption spectrometry using a flow-through electrochemical microcell

被引:36
作者
Bulska, E [1 ]
Walcerz, M [1 ]
Jedral, W [1 ]
Hulanicki, A [1 ]
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
关键词
cadmium; lead; FAAS; electrochemical preconcentration;
D O I
10.1016/S0003-2670(97)00537-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method was developed for the trace determination of lead and cadmium by flame atomic absorption spectrometry (FAAS) with electrochemical preconcentration. Both metals were deposited on the glassy-carbon electrode in a flow-through microcell of volume 3 mu l. At a consumption of a 3 mu I sample volume, the procedure resulted in an enhancement factor of more than 20 and yielded a detection limit of 1 mu g l(-1) for Pb and 0.08 mu g l(-1) for Cd. The relative standard deviation (n=12) was 2.1% at the 100 mu g l(-1) level for Pb and 3.2% at the 8 mu g l(-1) level for Cd. Results obtained for the standard reference material, water SRM 1643c, were in good agreement with the certified values. (C) 1997 Elsevier Science B.V.