Accuracy of physiologic deadspace measurement in intubated pediatric patients using a metabolic monitor: Comparison with the Douglas bag method

被引:19
作者
Lum, L
Saville, A
Venkataraman, ST
机构
[1] Univ Pittsburgh, Childrens Hosp Pittsburgh, Dept Anesthesiol CCM, Pittsburgh, PA 15213 USA
[2] Univ Pittsburgh, Childrens Hosp Pittsburgh, Dept Pediat, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Childrens Hosp Pittsburgh, Dept Respiratory Care, Pittsburgh, PA 15213 USA
关键词
physiologic deadspace; lung disease; children; Douglas bag; metabolic cart;
D O I
10.1097/00003246-199804000-00029
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the accuracy of physiologic deadspace (VD/VT) measurement, using a metabolic monitor. Design: prospective collection of data. Setting: University-affiliated children's hospital with a 51-bed critical cave area. Patients: infants and children who were sedated and paralyzed and were receiving mechanical ventilation through a cuffed endotracheal tube. Interventions: None. Measurements and Main Results: Mixed expired carbon dioxide tension (P (E) over bar co(2)) was measured. With the Douglas bag method, mixed expired gas was collected over 15 mins and was analyzed, With the metabolic monitor, F (E) over bar co(2) was measured for 15 mins and the results were averaged, The P (E) over bar co(2) was calculated by multiplying F (E) over bar co(2) by the barometric pressure. The Paco(2) was measured simultaneously. The P (E) over bar co(2) was corrected for the compressible Volume in the ventilator circuit, All gas volumes were corrected for body temperature, pressure, and water vapor pressure. The physiologic deadspace/tidal volume ratio (VD/VT) was calculated for both techniques using the Enghoff modification of the Bohr equation. The accuracy of the VD/VT measured, by using the metabolic monitor, was assessed by comparing this measurement against the VD/VT measured by the Douglas bag method. This comparison was done by simple linear regression and correlation and by bias analysis (Bland and Altman method), The magnitude of compressible volume expressed as a fraction of the tidal volume delivered by the ventilator was compared with the error in VD/VT expressed as the difference between the uncorrected and corrected VD/VT. Sixteen paired measurements were made in 12 children, The VD/VT measured by the metabolic monitor correlated well with the VD/VT measured by the Douglas bay method (r(2) = .99; p < .0001). There was no correlation between the bias of VD/VT and the average VD/VT. As the magnitude of compressible Volume increased, the error in VD/VT increased (r(2) = .61; p < .001). Conclusions: The VD/VT can be measured reliably and accurately in Intubated pediatric patients using a metabolic monitor. The metabolic monitor method is a convenient and simple alternative to the standard Douglas bag method.
引用
收藏
页码:760 / 764
页数:5
相关论文
共 14 条
[1]   RESPIRATORY DEADSPACE MEASUREMENTS IN NEONATES DURING EXTRACORPOREAL MEMBRANE-OXYGENATION [J].
ARNOLD, JH ;
THOMPSON, JE ;
BENJAMIN, PK .
CRITICAL CARE MEDICINE, 1993, 21 (12) :1895-1900
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
Bohr C., 1891, Skand Arch Physiol, V2, P236, DOI [DOI 10.1111/J.1748-1716.1891.TB00581.X, 10.1111/j.1748-1716.1891.tb00581.x]
[4]  
Enghoff Henrik, 1938, UPSALA LAKAREFOREN FORHANDL, V44, P191
[5]   THE CONCEPT OF DEADSPACE WITH SPECIAL REFERENCE TO THE SINGLE BREATH TEST FOR CARBON-DIOXIDE [J].
FLETCHER, R ;
JONSON, B ;
CUMMING, G ;
BREW, J .
BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (01) :77-88
[6]  
FLETCHER R, 1986, ANESTH ANALG, V65, P645
[7]  
JULIAN DG, 1960, J APPL PHYSIOL, V15, P37
[8]   DEAD SPACE VENTILATION IN NORMAL CHILDREN AND CHILDREN WITH OBSTRUCTIVE AIRWAYS DISEASE [J].
KERR, AA .
THORAX, 1976, 31 (01) :63-69
[9]   EVALUATION OF METABOLIC MEASURING-INSTRUMENTS FOR USE IN CRITICALLY ILL PATIENTS [J].
MAKITA, K ;
NUNN, JF ;
ROYSTON, B .
CRITICAL CARE MEDICINE, 1990, 18 (06) :638-644
[10]   INTERACTION OF SERIES AND PARALLEL DEAD SPACE IN THE LUNG [J].
PETRINI, MF ;
ROBERTSON, HT ;
HLASTALA, MP .
RESPIRATION PHYSIOLOGY, 1983, 54 (01) :121-136