Dose performance of a 64-channel dual-source CT scanner

被引:147
作者
McCollough, Cynthia H.
Primak, Andrew N.
Saba, Osama
Bruder, Herbert
Stierstorfer, Karl
Raupach, Rainer
Suess, Christoph
Schmidt, Bernhard
Ohnesorge, Bernd M.
Flohr, Thomas G.
机构
[1] Mayo Clin & Mayo Fdn, Coll Med, Dept Radiol, CT Clin Innovat Ctr, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA USA
[3] Siemens Med Solut, Forchheim, Germany
[4] Univ Tubingen, Dept Diagnost Radiol, Tubingen, Germany
关键词
D O I
10.1148/radiol.2433061165
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively compare the dose performance of a 64-channel multi-detector row computed tomographic ( CT) scanner and a 64-channel dual-source CT scanner from the same manufacturer. Materials and Methods: To minimize dose in the cardiac ( dual-source) mode, the evaluated dual-source CT system uses a cardiac beam-shaping filter, three-dimensional adaptive noise reduction, heart rate-dependent pitch, and electrocardiographically based modulation of the tube current. Weighted CT dose index per 100 mAs was measured for the head, body, and cardiac beam-shaping filters. Kerma-length product was measured in the spiral cardiac mode at four pitch values and three electrocardiographic modulation temporal windows. Noise was measured in an anthropomorphic phantom. Data were compared with data from a 64-channel multi-detector row CT scanner. Results: For the multi-detector row and dual-source CT systems, respectively, weighted CT dose index per 100 mAs was 14.2 and 12.2 mGy ( head CT), 6.8 and 6.4 mGy ( body CT), and 6.8 and 5.3 mGy ( cardiac CT). In the spiral cardiac mode ( no electrocardiographically based tube current modulation, 0.2 pitch), equivalent noise occurred at volume CT dose index values of 23.7 and 35.0 mGy ( coronary artery calcium CT) and 58.9 and 61.2 mGy ( coronary CT angiography) for multi-detector row CT and dual-source CT, respectively. The use of heart rate-dependent pitch values reduced volume CT dose index to 46.2 mGy ( 0.265 pitch), 34.0 mGy ( 0.36 pitch), and 26.6 mGy ( 0.46 pitch) compared with 61.2 mGy for 0.2 pitch. The use of electrocardiographically based tube current-modulation and temporal windows of 110, 210, and 310 msec further reduced volume CT dose index to 9.1-25.1 mGy, dependent on the heart rate. Conclusion: For electrocardiographically gated coronary CT angiography, image noise equivalent to that of multi-detector row CT can be achieved with dual-source CT at doses comparable to or up to a factor of two lower than the doses at multi-detector row CT, depending on heart rate of the patient. (c) RSNA, 2007.
引用
收藏
页码:775 / 784
页数:10
相关论文
共 29 条
[1]   Contrast-enhanced coronary artery visualization by dual-source computed tomography - Initial experience [J].
Achenbach, S ;
Ropers, D ;
Kuettner, A ;
Flohr, T ;
Ohnesorge, B ;
Bruder, H ;
Theessen, H ;
Karakaya, M ;
Daniel, WG ;
Bautz, W ;
Kalender, WA ;
Anders, K .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) :331-335
[2]  
*EUR COMM, 2000, EUR GUID QUAL CRIT C
[3]   Image reconstruction and performance evaluation for ECG-gated spiral scanning with a 16-slice CT system [J].
Flohr, T ;
Ohnesorge, B ;
Bruder, H ;
Stierstorfer, K ;
Simon, J ;
Suess, C ;
Schaller, S .
MEDICAL PHYSICS, 2003, 30 (10) :2650-2662
[4]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[5]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[6]   Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot [J].
Flohr, TG ;
Stierstorfer, K ;
Ulzheimer, S ;
Bruder, H ;
Primak, AN ;
McCollough, CH .
MEDICAL PHYSICS, 2005, 32 (08) :2536-2547
[7]   Advances in cardiac imaging with 16-section CT systems [J].
Flohr, TG ;
Schoepf, UJ ;
Kuettner, A ;
Halliburton, S ;
Bruder, H ;
Suess, C ;
Schmidt, B ;
Hofmann, L ;
Yucel, EK ;
Schaller, S ;
Ohnesorge, BM .
ACADEMIC RADIOLOGY, 2003, 10 (04) :386-401
[8]  
Halliburton SS, 2003, HERZ, V28, P20, DOI 10.1007/s00059-003-2445-4
[9]  
ICRP (International Commission on Radiological Protection), 1991, ANN ICRP, V21, P1990
[10]  
Interactive Harris., 2002, PRIVACY AM BUSINESS, P1