Efficiency calibration and coincidence summing correction for a 4πNaI(Tl detector array

被引:30
作者
Byun, SH [1 ]
Prestwich, WV [1 ]
Chin, K [1 ]
McNeill, FE [1 ]
Chettle, DR [1 ]
机构
[1] McMaster Univ, Med Phys & Appl Radiat Sci Unit, Hamilton, ON L8S 4K1, Canada
关键词
NaI(Tl) detector; efficiency; coincidence summing; Monte Carlo simulation; in vivo neutron activation;
D O I
10.1016/j.nima.2004.06.174
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
Efficiency calibration and coincidence summing correction were performed for an array of NaI(Tl) detectors developed for in vivo neutron activation analysis. The detector array consisted of eight large NaI(Tl)s with square cross-section and the enclosing geometry of the detection produced both extremely high efficiency and a severe coincidence summing effect. The efficiency was calibrated mainly by Monte Carlo simulation with the EGS4 code. In order to obtain the coincidence summing correction factors through the simulation the coincident emission of photons was modeled. The influence of the angular correlation was explicitly investigated by adding the angle-correlated sampling routine in the simulation code. The experimental efficiencies and correction factors were measured using Cs-137, Co-60 and Na-22. For both peak and total efficiencies, experiment and simulation typically agreed within 5%. The discrepancy between simulation and experiment for the coincidence summing correction factors was 15% or less. The approximate methods adopted in the measurement were analyzed in detail and estimated to be reliable for calibrating high-efficiency detectors. The calculation was finally expanded from point sources to realistic volume sources. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:674 / 685
页数:12
相关论文
共 30 条
[1]
Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computation [J].
Abbas, MI .
APPLIED RADIATION AND ISOTOPES, 2001, 55 (02) :245-252
[2]
An accelerator based system for in vivo neutron activation analysis measurements of manganese in human hand bones [J].
Arnold, ML ;
McNeill, FE ;
Stronach, IM ;
Pejovic-Milic, A ;
Chettle, DR ;
Waker, A .
MEDICAL PHYSICS, 2002, 29 (11) :2718-2724
[3]
Calibration of the well-type germanium gamma-ray detector employing two gamma-ray spectra [J].
Blaauw, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 419 (01) :146-153
[4]
The use of sources emitting coincident gamma-rays for the determination of photopeak and total efficiencies in non-point-source counting geometries [J].
Blaauw, M ;
Ammerlaan, MJJ ;
Gelsema, SJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 385 (02) :330-334
[5]
*BNL, EV NUCL STRUCT DAT F
[6]
ABSOLUTE STANDARDIZATION WITH A NAI (T1) CRYSTAL .3. CALIBRATION OF BETA+-EMITTERS [J].
BRINKMAN, GA ;
ATEN, AHW .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1963, 14 (10) :503-&
[7]
Monte Carlo simulation of the response of a large NaI(Tl) total absorption spectrometer for β-decay studies [J].
Cano-Ott, D ;
Tain, JL ;
Gadea, A ;
Rubio, B ;
Batist, L ;
Karny, M ;
Roeckl, E .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 430 (2-3) :333-347
[8]
COINCIDENCE SUMMING CORRECTIONS IN GE(LI)-SPECTROMETRY AT LOW SOURCE-TO-DETECTOR DISTANCES [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1979, 158 (2-3) :471-477
[9]
Debertin K., 1988, Gammaand X-Ray Spectrometry with Semiconductor Detectors
[10]
INVIVO MONITORING OF SKELETAL ALUMINUM BURDEN IN PATIENTS WITH RENAL-FAILURE [J].
ELLIS, KJ ;
KELLEHER, S ;
RACITI, A ;
SAVORY, J ;
WILLS, M .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1988, 124 (01) :85-95