Efficient scatter distribution estimation and correction in CBCT using concurrent Monte Carlo fitting

被引:52
作者
Bootsma, G. J. [1 ]
Verhaegen, F. [2 ,3 ]
Jaffray, D. A. [1 ,4 ,5 ]
机构
[1] Princess Margaret Canc Ctr, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Maastricht Univ, Med Ctr, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, NL-6201 BN Maastricht, Netherlands
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
[4] Princess Margaret Canc Ctr, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[5] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
关键词
x-ray scatter; cone-beam CT; Monte Carlo; image quality; X-RAY SCATTER; LOCALLY WEIGHTED REGRESSION; PRIMARY MODULATION; BEAM; RADIATION; SIMULATIONS; REDUCTION; GEOMETRY; IMAGES;
D O I
10.1118/1.4903260
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: X-ray scatter is a significant impediment to image quality improvements in cone-beam CT (CBCT). The authors present and demonstrate a novel scatter correction algorithm using a scatter estimation method that simultaneously combines multiple Monte Carlo (MC) CBCT simulations through the use of a concurrently evaluated fitting function, referred to as concurrent MC fitting (CMCF). Methods: The CMCF method uses concurrently run MC CBCT scatter projection simulations that are a subset of the projection angles used in the projection set, P, to be corrected. The scattered photons reaching the detector in each MC simulation are simultaneously aggregated by an algorithm which computes the scatter detector response, S-MC. S-MC is fit to a function, S-F, and if the fit of S-F is within a specified goodness of fit (GOF), the simulations are terminated. The fit, S-F, is then used to interpolate the scatter distribution over all pixel locations for every projection angle in the set P. The CMCF algorithm was tested using a frequency limited sum of sines and cosines as the fitting function on both simulated and measured data. The simulated data consisted of an anthropomorphic head and a pelvis phantom created from CT data, simulated with and without the use of a compensator. The measured data were a pelvis scan of a phantom and patient taken on an Elekta Synergy platform. The simulated data were used to evaluate various GOF metrics as well as determine a suitable fitness value. The simulated data were also used to quantitatively evaluate the image quality improvements provided by the CMCF method. A qualitative analysis was performed on the measured data by comparing the CMCF scatter corrected reconstruction to the original uncorrected and corrected by a constant scatter correction reconstruction, as well as a reconstruction created using a set of projections taken with a small cone angle. Results: Pearson's correlation, r, proved to be a suitable GOF metric with strong correlation with the actual error of the scatter fit, S-F. Fitting the scatter distribution to a limited sum of sine and cosine functions using a low-pass filtered fast Fourier transform provided a computationally efficient and accurate fit. The CMCF algorithm reduces the number of photon histories required by over four orders of magnitude. The simulated experiments showed that using a compensator reduced the computational time by a factor between 1.5 and 1.75. The scatter estimates for the simulated and measured data were computed between 35-93 s and 114-122 s, respectively, using 16 Intel Xeon cores (3.0 GHz). The CMCF scatter correction improved the contrast-to-noise ratio by 10%-50% and reduced the reconstruction error to under 3% for the simulated phantoms. Conclusions: The novel CMCF algorithm significantly reduces the computation time required to estimate the scatter distribution by reducing the statistical noise in the MC scatter estimate and limiting the number of projection angles that must be simulated. Using the scatter estimate provided by the CMCF algorithm to correct both simulated and real projection data showed improved reconstruction image quality. (c) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:54 / 68
页数:15
相关论文
共 40 条
[1]
[Anonymous], PIRS701
[2]
Accelerating Monte Carlo simulations of photon transport in a voxelized geometry using a massively parallel graphics processing unit [J].
Badal, Andreu ;
Badano, Aldo .
MEDICAL PHYSICS, 2009, 36 (11) :4878-4880
[3]
Correction of CT artifacts and its influence on Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Beaulieu, Luc ;
Palefsky, Steven ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2007, 34 (06) :2119-2132
[4]
MONTE-CARLO SIMULATION OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :713-720
[5]
Spatial frequency spectrum of the x-ray scatter distribution in CBCT projections [J].
Bootsma, G. J. ;
Verhaegen, F. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2013, 40 (11)
[6]
The effects of compensator and imaging geometry on the distribution of x-ray scatter in CBCT [J].
Bootsma, G. J. ;
Verhaegen, F. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :897-914
[7]
Bootsma G. J., 2011, P SOC PHOTO-OPT INS, V7961
[8]
Segmenting the prostate and rectum in CT imagery using anatomical constraints [J].
Chen, Siqi ;
Lovelock, D. Michael ;
Radke, Richard J. .
MEDICAL IMAGE ANALYSIS, 2011, 15 (01) :1-11
[9]
An innovative phantom for quantitative and qualitative investigation of advanced x-ray imaging technologies [J].
Chiarot, CB ;
Siewerdsen, JH ;
Haycocks, T ;
Moseley, DJ ;
Jaffray, DA .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21) :N287-N297
[10]