Dose reduction in gastrointestinal and genitourinary fluoroscopy: Use of grid-controlled pulsed fluoroscopy

被引:51
作者
Boland, GWL [1 ]
Murphy, B [1 ]
Arellano, R [1 ]
Niklason, L [1 ]
Mueller, PR [1 ]
机构
[1] Massachusetts Gen Hosp, Div Adbominal Imaging & Intervent Radiol, Boston, MA 02114 USA
关键词
D O I
10.2214/ajr.175.5.1751453
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE, We evaluated the diagnostic accuracy of a grid-controlled fluoroscopy unit compared with a conventional continuous fluoroscopy unit for a variety of abdominal and pelvic fluoroscopic examinations. SUBJECTS AND METHODS. Seventy patients (29 men and 41 women; age range, 24-78 years) were enrolled in one of seven abdominal and pelvic fluoroscopic examinations, including upper gastrointestinal series (n = 20), barium enema (n = 10), voiding cystourethrogram (n = 10), percutaneous abdominal catheter tube injection (n = 10), hysterosalpingogram (n = 10), and percutaneous needle insertion and catheter placement (nephrostomy, percutaneous biliary drainage) (n = 10). Each patient underwent at least 10 sec of continuous fluoroscopy that was randomly and blindly compared with 10-sec periods of pulsed fluoroscopy at 15, 7.5, and 3.75 frames per second. A radiologist outside the examination room, unaware of the frame rate per second, evaluated the procedure in real time on a television monitor. The radiologist assessed image quality and diagnostic acceptability using a scoring system. Statistical analysis was performed using the paired Student's t test. RESULTS. For all procedures at all frame rates, we found no statistically significant superiority of one frame rate over another. For most procedures, the slower frame rates were considered equivalent to continuous fluoroscopy when the images were assessed for image quality and diagnostic confidence. CONCLUSION. Our findings suggest that most abdominal and pelvic fluoroscopic procedures can be performed at substantially lower frame rates than those used for continuous fluoroscopy; adopting this procedure may lead to substantial dose savings for the patient and the fluoroscopy operator.
引用
收藏
页码:1453 / 1457
页数:5
相关论文
共 10 条
[1]  
AMMANN E, 1995, CATEGORICAL COURSE P, P59
[2]   PERCEPTUAL COMPARISON OF PULSED AND CONTINUOUS FLUOROSCOPY [J].
AUFRICHTIG, R ;
XUE, P ;
THOMAS, CW ;
GILMORE, GC ;
WILSON, DL .
MEDICAL PHYSICS, 1994, 21 (02) :245-256
[3]  
BERTHELSEN B, 1991, ACTA RADIOL, V32, P492
[4]   DOSE REDUCTION LOW PULSE-RATE FLUOROSCOPY [J].
GROLLMAN, JH ;
KLOSTERMAN, H ;
MOLER, CL ;
HERMAN, MW ;
EBER, LM ;
MACALPIN, RN .
RADIOLOGY, 1972, 105 (02) :293-+
[5]   RADIATION REDUCTION BY MEANS OF LOW PULSE-RATE FLUOROSCOPY DURING CARDIAC-CATHETERIZATION AND CORONARY ARTERIOGRAPHY [J].
GROLLMAN, JH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1974, 121 (03) :636-641
[6]  
Hahn H., 1997, Medicamundi, V41, P12
[7]   EFFECT OF PULSED PROGRESSIVE FLUOROSCOPY ON REDUCTION OF RADIATION-DOSE IN THE CARDIAC-CATHETERIZATION LABORATORY [J].
HOLMES, DR ;
WONDROW, MA ;
GRAY, JE ;
VETTER, RJ ;
FELLOWS, JL ;
JULSRUD, PR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (01) :159-162
[8]  
International Commission on Radiation Protection, 1977, 26 INT COMM RAD PROT
[9]   RADIATION EXPOSURE TO PATIENTS AND MEDICAL PERSONNEL DURING RADIOFREQUENCY CATHETER ABLATION FOR SUPRAVENTRICULAR TACHYCARDIA [J].
LINDSAY, BD ;
EICHLING, JO ;
AMBOS, HD ;
CAIN, ME .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (02) :218-223
[10]  
*NAT COUNC RAD PRO, 1989, 100 INT COMM RAD PRO