A hip joint simulator study of the performance of metal-on-metal joints - Part I: The role of materials

被引:101
作者
Dowson, D
Hardaker, C
Flett, M
Isaac, GH
机构
[1] DePuy Int Ltd, Leeds LS11 8DT, W Yorkshire, England
[2] Univ Leeds, Sch Mech Engn, Leeds, W Yorkshire, England
关键词
metal-on-metal hip arthroplasties; resurfacing; simulator; materials composition; processing; running-in water;
D O I
10.1016/j.arth.2004.09.015
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The use of metal-on-metal low-wearing bearings has promoted great interest in the factors determining the volume of wear debris generated in such joints, including the developing surface replacement alternatives. Twenty-six pairs of low- and high-carbon components in wrought or cast form of 36 rum nominal diameter and exhibiting similar clearances were Studied over 5 million test cycles in a 10-station hip joint simulator. Low-carbon cast materials exhibited higher wear than high-carbon cast or wrought materials. Little difference was found in the running-in wear volumes generated by high-carbon wrought or cast materials, but the wrought material exhibited a slight advantage at the smallest diametral clearance considered of 105 mum. Volumetric wear appeared to decrease as the diametral clearance decreased. Key words: metal-on-metal hip arthroplasties, resurfacing, simulator, materials composition, processing, running-in wear. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:118 / 123
页数:6
相关论文
共 9 条
[1]  
CHAN FW, 1997, ARCH AM ACAD ORTHOPA, V1, P57
[2]   SURVIVORSHIP OF TOTAL HIP REPLACEMENTS [J].
DOBBS, HS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1980, 62 (02) :168-173
[3]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130
[4]  
DOWSON D, 2003, P 29 LEEDS LYON S TR, V41, P273
[5]   Development of a ten-station, multi-axis hip joint simulator [J].
Goldsmith, AAJ ;
Dowson, D .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H4) :311-316
[6]  
MEDLEY JB, 1996, CLIN ORTHOP S, V329, P148
[7]   INVESTIGATION OF THE TRIBOLOGICAL BEHAVIOR OF METAL-ON-METAL COMBINATIONS FOR ARTIFICIAL HIP JOINTS [J].
STREICHER, RM ;
SCHON, R ;
SEMLITSCH, MF .
BIOMEDIZINISCHE TECHNIK, 1990, 35 (05) :107-111
[8]   TRIBOLOGY (FRICTION, LUBRICATION AND WEAR) OF ALL-METAL ARTIFICIAL HIP JOINTS [J].
WALKER, PS ;
GOLD, BL .
WEAR, 1971, 17 (04) :285-&
[9]   COMPARATIVE STUDY OF TOTAL HIP-REPLACEMENT PROSTHESES [J].
WEIGHTMAN, BO ;
PAUL, IL ;
ROSE, RM ;
SIMON, SR ;
RADIN, EL .
JOURNAL OF BIOMECHANICS, 1973, 6 (03) :299-+