Effects of hydroxyapatite coating on Ti-6Al-4V implant-site infection in a rabbit tibial model

被引:41
作者
Vogely, HC
Oosterbos, CJM
Puts, EWA
Nijhof, MW
Nikkels, PGJ
Fleer, A
Tonino, AJ
Dhert, WJA
Verbout, AJ
机构
[1] Univ Utrecht, Med Ctr, Dept Orthopaed, NL-3508 GA Utrecht, Netherlands
[2] Univ Utrecht, Med Ctr, Dept Pathol, NL-3508 GA Utrecht, Netherlands
[3] Univ Utrecht, Med Ctr, Dept Med Microbiol, NL-3508 GA Utrecht, Netherlands
[4] Ziekenhuis Zeeuws Vlaanderen, Dept Orthopaed, Terneuzen, Netherlands
[5] Atrium Med Ctr, Dept Orthopaed, Heerlen, Netherlands
关键词
D O I
10.1002/jor.1100180323
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
To investigate the effect of implant type after direct contamination, a hydroxyapatite-coated or noncoated Ti-6A1-4V implant was inserted into both tibiae of 32 New Zealand White rabbits. Prior to implantation, the left tibia was contaminated with increasing concentrations of Staphylococcus aureus (10(2)-10(5) colony-forming units), ranging from very low (10(2)) to relatively high (10(5)). Four weeks after implantation, half of the tibial bone adjacent to the implant was harvested for bacteriological examination. Bacterial counts were quantified by plating serial dilutions. For the histological evaluation, sections of the implant with the remaining tibia were examined by semiquantitative scoring of infection parameters. The bacteriological data showed the inoculum dose and implant type to have a significant effect on the culture outcome: more bacteria were retrieved from the hydroxyapatite-coated implants than from the noncoated titanium implants. Histological evaluation showed an increased score for the infected left tibiae compared with their contralateral control. In addition, with increasing inoculum dosage, the difference between the two types of implant increased. We demonstrated that infections can occur with biocompatible, noncemented implants and that they are related to the dose of the original inoculum. Bacteria were more likely to grow onto or next to the hydroxyapatite implants than on titanium implants and resulted in a more severe histopathological characterization of infection.
引用
收藏
页码:485 / 493
页数:9
相关论文
共 33 条
[1]   New model of experimental prosthetic joint infection due to methicillin-resistant Staphylococcus aureus: A microbiologic, histopathologic, and magnetic resonance imaging characterization [J].
Belmatoug, N ;
Cremieux, AC ;
Bleton, R ;
Volk, A ;
SalehMghir, A ;
Grossin, M ;
Garry, L ;
Carbon, C .
JOURNAL OF INFECTIOUS DISEASES, 1996, 174 (02) :414-417
[2]  
Blomgren G, 1981, Acta Orthop Scand Suppl, V187, P1
[3]   INFLUENCE OF METAL IMPLANTS ON INFECTION - AN EXPERIMENTAL-STUDY IN RABBITS [J].
CORDERO, J ;
MUNUERA, L ;
FOLGUEIRA, MD .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :717-720
[4]  
Cordero J, 1996, INJURY, V27, P34
[5]   Influence of bacterial strains on bone infection [J].
Cordero, J ;
Munuera, L ;
Folgueira, MD .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1996, 14 (04) :663-667
[6]   ROLE OF FIBRONECTIN IN STAPHYLOCOCCAL ADHESION TO METALLIC SURFACES USED AS MODELS OF ORTHOPEDIC DEVICES [J].
DELMI, M ;
VAUDAUX, P ;
LEW, DP ;
VASEY, H .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (03) :432-438
[7]   A HISTOLOGICAL AND HISTOMORPHOMETRICAL INVESTIGATION OF FLUORAPATITE, MAGNESIUMWHITLOCKITE, AND HYDROXYLAPATITE PLASMA-SPRAYED COATINGS IN GOATS [J].
DHERT, WJA ;
KLEIN, CPAT ;
JANSEN, JA ;
VANDERVELDE, EA ;
VRIESDE, RC ;
ROZING, PM ;
DEGROOT, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (01) :127-138
[8]   A MECHANICAL INVESTIGATION OF FLUORAPATITE, MAGNESIUMWHITLOCKITE, AND HYDROXYLAPATITE PLASMA-SPRAYED COATINGS IN GOATS [J].
DHERT, WJA ;
KLEIN, CPAT ;
WOLKE, JGC ;
VANDERVELDE, EA ;
DEGROOT, K ;
ROZING, PM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (10) :1183-1200
[9]  
Dhert WJA, 1998, J BIOMED MATER RES, V41, P574, DOI 10.1002/(SICI)1097-4636(19980915)41:4<574::AID-JBM9>3.0.CO
[10]  
2-9