Evaluation of basic algorithms for transferring quantitative multivariate calibrations between scanning grating and FT NIR spectrometers

被引:32
作者
Leion, H [1 ]
Folestad, S [1 ]
Josefson, M [1 ]
Sparén, A [1 ]
机构
[1] AstraZeneca R&D, Analyt Dev, Pharmaceut & R&D, SE-43183 Molndal, Sweden
关键词
near infrared spectroscopy; diffuse-reflectance; calibration transfer; multivariate quantitative calibrations; partial least squares regression; solid pharmaceutical formulations;
D O I
10.1016/j.jpba.2004.09.046
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A key issue in near infrared spectroscopy (NIR) is the possibility to use calibrations generated on one instrument for predictions on others. A number of methods for calibration transfer have been proposed, but which method to choose is typically not straightforward. An evaluation of a number of methods for transferring quantitative calibrations between different instruments was carried out on near infrared diffuse-reflectance data from a pharmaceutical formulation. Six instruments were included in the study, five of which were scanning grating instruments, both with and without fibre-optic probe configuration, and one of which was a Fourier-transform instrument, equipped with a fibre-optic probe. The results show that it is possible to transfer calibrations between different instruments, provided that a structured procedure is used. Simple techniques for calibration transfer, such as slope/bias correction on the predicted results, as well as standard normal variate transformation and local centring of the raw spectra, gave considerably lower prediction errors on transfer than did standardisation with a certified diffuse-reflectance standard, or direct transfer without any transfer function. Notably, including more than one instrument in the calibration also improved the prediction ability of the models on calibration transfer. No significant differences in wavelength scale were found when a certified diffuse-reflectance wavelength standard was measured on the instruments studied. Nor did simulated wavelength scale differences below +/-0.3 nm cause any significant change in the prediction errors. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:47 / 55
页数:9
相关论文
共 52 条
  • [1] *ABB BOM INC, MB SER TFIR SPEC FEA
  • [2] Quantitative analysis of film coating in a fluidized bed process by in line NIR spectrometry and multivariate batch calibration
    Andersson, M
    Folestad, S
    Gottfries, J
    Johansson, MO
    Josefson, M
    Wahlund, KG
    [J]. ANALYTICAL CHEMISTRY, 2000, 72 (09) : 2099 - 2108
  • [3] Monitoring of a film coating process for tablets using near infrared reflectance spectrometry
    Andersson, M
    Josefson, M
    Langkilde, FW
    Wahlund, KG
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 20 (1-2) : 27 - 37
  • [4] [Anonymous], [No title captured]
  • [5] STANDARD NORMAL VARIATE TRANSFORMATION AND DE-TRENDING OF NEAR-INFRARED DIFFUSE REFLECTANCE SPECTRA
    BARNES, RJ
    DHANOA, MS
    LISTER, SJ
    [J]. APPLIED SPECTROSCOPY, 1989, 43 (05) : 772 - 777
  • [6] Quantitative determination of content in binary powder mixtures using diffuse reflectance near infrared spectrometry and multivariate analysis
    Berntsson, O
    Danielsson, LG
    Johansson, MO
    Folestad, S
    [J]. ANALYTICA CHIMICA ACTA, 2000, 419 (01) : 45 - 54
  • [7] Estimation of effective sample size when analysing powders with diffuse reflectance near-infrared spectrometry
    Berntsson, O
    Danielsson, LG
    Folestad, S
    [J]. ANALYTICA CHIMICA ACTA, 1998, 364 (1-3) : 243 - 251
  • [8] Development and validation of methods for the determination of miokamycin in various pharmaceutical preparations by use of near infrared reflectance spectroscopy
    Blanco, M
    Coello, J
    Eustaquio, A
    Itturriaga, H
    Maspoch, S
    [J]. ANALYST, 1999, 124 (07) : 1089 - 1092
  • [9] Near-infrared spectroscopy in the pharmaceutical industry
    Blanco, M
    Coello, J
    Iturriaga, H
    Maspoch, S
    de la Pezuela, C
    [J]. ANALYST, 1998, 123 (08) : 135R - 150R
  • [10] Transfer of near-infrared multivariate calibrations without standards
    Blank, TB
    Sum, ST
    Brown, SD
    Monfre, SL
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (17) : 2987 - 2995