Serial measurement of polyethylene wear of well-fixed cementless metal-backed acetabular component in total hip arthroplasty: An over 10 year follow-up study

被引:16
作者
Dai, XS [1 ]
Omori, H [1 ]
Okumura, Y [1 ]
Ando, M [1 ]
Oki, H [1 ]
Hashimoto, N [1 ]
Baba, H [1 ]
机构
[1] Fukui Med Univ, Sch Med, Dept Orthoped Surg, Fukui 9101193, Japan
关键词
total hip arthroplasty; polyethylene-acetabular component; wear;
D O I
10.1046/j.1525-1594.2000.06571-2.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Serial radiographic measurements of polyethylene wear were performed in 38 hips (33 patients) with primary cementless total hip arthroplasty (THA). The average follow-up period was 131.8 months. All prostheses were assessed as radiographically stable at the latest follow-up. A two-dimensional method was used to calculate the relative migration of the femoral head center to the cup center. The average total linear wear and wear rate were 1.22 mm and 0.11 mm/year, respectively. The degree of wear in the first 2 postoperative years accounted for nearly 40% of the total wear at the end of the study (average follow up: 131.8 +/- 10.0 months, +/-SD). The migration of the femoral head at an average period of 3.4 months after operation accounted for 56% of the amount of wear in the first 2 years. Wear rate decreased gradually with time and stabilized after the fourth year. However, in 2 patients, a progressive increase in the wear rate was associated with severe osteolysis and failure of THA. Both creep and wear contributed to the femoral penetration into the polyethylene liner. The influence of creep cannot be ruled out, especially in the early period after operation. Polyethylene wear is a multifactorial process, and the study of individual wear patterns might be useful in identifying patients who are at risk of late failure of THA.
引用
收藏
页码:746 / 751
页数:6
相关论文
共 24 条
[1]  
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[2]  
[Anonymous], CLIN ORTHOPAEDICS
[3]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[4]  
BROWN KJ, 1975, PLASTICS MED SURG
[5]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[6]  
Devane PA, 1995, CLIN ORTHOP RELAT R, V319, P303
[7]  
Dorr LD, 1998, CLIN ORTHOP RELAT R, P158
[8]  
Engh CA., 1990, CLIN ORTHOP RELAT R, V257, P107
[9]  
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46
[10]  
HODGKINSON JP, 1988, CLIN ORTHOP RELAT R, V228, P105