The Self-adjusting File (SAF). Part 2: Mechanical Analysis

被引:61
作者
Hof, Rafael [1 ]
Perevalov, Valery [1 ]
Eltanani, Moshe [1 ]
Zary, Raviv [1 ]
Metzger, Zvi [1 ,2 ]
机构
[1] ReDent Nova Inc, Raanana, Israel
[2] Tel Aviv Univ, Maurice & Gabriela Goldschleger Sch Dent Med, Dept Endodontol, IL-69978 Tel Aviv, Israel
关键词
Cleaning and shaping; cyclic fatigue; endodontic files; fatigue; mechanical properties; SAF; self-adjusting file; NICKEL-TITANIUM INSTRUMENTS; ROTARY INSTRUMENTS; SHAPING ABILITY; EXTRACTED TEETH; ROOT CANALS; OVAL CANALS; PROTAPER; SYSTEMS; PROFILE; RACE;
D O I
10.1016/j.joen.2009.12.028
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The study was designed to explore the mechanical properties of the self-adjusting file (SAF) and its application in the root canal using continuous irrigation. Methods: The compressibility of the SAF file and the resulting peripheral force were measured using specially designed systems. The abrasivity of the file was tested on dentin blocks representing a flat root canal. The durability of the SAF file was tested using a functional fatigue-to-failure assay. Degradation of the file was evaluated by using files that were previously used for 10, 20, and 30 minutes and comparing their efficacy with that of new, unused files. The potential of extruding irrigant beyond the apex was explored in roots with an open apical foramen. Results: The SAF file was elastically compressible from a diameter of 1.5 mm to dimensions similar to those of a #20 stainless steel K-file. This compression resulted in an evenly applied force to the root canal walls. The in-and-out vibration of the file and the peripheral force, combined with its abrasivity, allow for hard-tissue removal. Under the conditions of the experiment, no mechanical failure was observed with up to 29 minutes of operation in the root canal. The file loses its efficacy after prolonged use, with a 40% reduction after 30 minutes of operation. The operation of the file with continuous irrigation did not push the irrigant beyond an open apical foramen. Conclusions: The SAF file is an elastically compressible file that effectively removes dentin and can mechanically endure use under its recommended mode of operation with a minimal loss of efficacy. (J Endod 2010;36:691-696)
引用
收藏
页码:691 / 696
页数:6
相关论文
共 16 条
[1]  
*AM DENT ASS, 2006, ADA PROFESSIONAL PRO, V1, P11
[2]   Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: Value of the safety quotient [J].
Blum, JY ;
Machtou, P ;
Ruddle, C ;
Micallef, JP .
JOURNAL OF ENDODONTICS, 2003, 29 (09) :567-575
[3]   A laboratory analysis of gutta-percha-filled area obtained using Thermafil, System B and lateral condensation [J].
De-Deus, G ;
Gurgel, ED ;
Magalhaes, KM ;
Coutinho, T .
INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (05) :378-383
[4]   A COMPARISON OF CANAL PREPARATION WITH NICKEL-TITANIUM AND STAINLESS-STEEL INSTRUMENTS [J].
ESPOSITO, PT ;
CUNNINGHAM, CJ .
JOURNAL OF ENDODONTICS, 1995, 21 (04) :173-176
[5]   A comparison of Profile, Hero 642, and K3 instrumentation systems in teeth using digital imaging analysis [J].
González-Rodríguez, MP ;
Ferrer-Luque, CM .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2004, 97 (01) :112-115
[6]  
Hulsmann M., 2005, Endodontic Topics, V10, P30, DOI DOI 10.1111/J.1601-1546.2005.00152.X
[7]  
*ISO, 2008, ISO36301
[8]   Comparison between a Novel Nickel-Titanium Alloy and 508 Nitinol on the Cyclic Fatigue Life of ProFile 25/.04 Rotary Instruments [J].
Johnson, Eugenia ;
Lloyd, Adain ;
Kuttler, Sergio ;
Namerow, Kenneth .
JOURNAL OF ENDODONTICS, 2008, 34 (11) :1406-1409
[9]  
METZGER Z, 2010, COHENS PATHWAYS PULP
[10]  
METZGER Z, J ENDOD IN PRESS