Rupture of PTCA balloons - an in vitro examination

被引:1
作者
Dubel, HP
Gliech, V
Bunk, G
Kubat, B
Schneider, M
Kleber, FX
机构
[1] Humboldt Univ, Klinikum Charite, Fak Med,Zentrum Innere Med, Med Klin & Poliklin 1,Herzkatheterbereich, D-10098 Berlin, Germany
[2] Inst Med Informat & Biometrie, D-10098 Berlin, Germany
来源
ZEITSCHRIFT FUR KARDIOLOGIE | 1997年 / 86卷 / 12期
关键词
PTCA; balloon-rupture; burst pressure; compliance; in vitro measurement;
D O I
10.1007/s003920050138
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A large number of PTCA balloon catheters from different manufacturers were examined following their use. Several of these balloons had been previously used, for PTCA for the mere purpose of guide wire support. Each balloon was evaluated starting at 8 atm. The pressure was increased in steps of 1 atm until rupture. This stepwise increase of pressure was documented at each Les el in a short cine film. During dilatations, balloons were in room air until rupture. None of the balloons rupture il before the rated burst pressure. A few or the balloons tolerated pressures of approximately 27 aim. With increasing Pressure the shape of the balloons became distorted. Most assumed a banana shape. Most of the balloons ruptured after developing a small hole. Increasing pressure produced increasing diameter and, generally, we found a linear increase. Most balloons increased by > 15 % in diameter over the nominated diameter before: rupture. The burst pressure in all balloons was not influenced by the prior mode of dilation, duration of dilation, maximum pressure applied, or length of the balloon. Balloons with smaller diameter ruptured later than compared to those with larger diameters. Our findings demonstrated that all of the balloons tolerated the manufacturers' recommended inflation pressure supporting the safety in their use.
引用
收藏
页码:968 / 974
页数:17
相关论文
共 27 条
[2]   BALLOON CATHETER CORONARY ANGIOPLASTY - BALLOON RUPTURE [J].
ANGELINI, P .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 20 (02) :150-151
[3]   PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - A REVIEW OF CURRENT BALLOON DILATATION SYSTEMS [J].
AVEDISSIAN, MG ;
KILLEAVY, ES ;
GARCIA, JM ;
DEAR, WE .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 18 (04) :263-275
[4]   CORONARY DISSECTION RESULTING FROM ANGIOPLASTY WITH SLOW OSCILLATING VS RAPID INFLATION AND SLOW VS RAPID DEFLATION [J].
BLANKENSHIP, JC ;
FORD, AC ;
HENRY, SD ;
FREY, CM .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1995, 34 (03) :202-209
[5]   INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE [J].
COLOMBO, A ;
HALL, P ;
NAKAMURA, S ;
ALMAGOR, Y ;
MAIELLO, L ;
MARTINI, G ;
GAGLIONE, A ;
GOLDBERG, SL ;
TOBIS, JM .
CIRCULATION, 1995, 91 (06) :1676-1688
[6]  
GERBER TL, 1988, CIRCULATION S1, V88, P587
[7]   EFFECTS OF BALLOON COMPLIANCE ON ANGIOGRAPHIC AND CLINICAL OUTCOMES AFTER PTCA [J].
GOEL, P ;
AGARWAL, R ;
KAUL, U ;
WASIR, HS .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1995, 51 (01) :1-4
[8]   ASSESSMENT OF ANGIOPLASTY BALLOON CATHETERS - PRELIMINARY STUDIES [J].
HENSON, JHL ;
JEANS, WD ;
NEWHAM, FJ ;
PAICE, C ;
BLAKEBOROUGH, A .
BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (731) :1026-1031
[9]  
Hering D, 1996, Z KARDIOL, V85, P273
[10]   METHODOLOGICAL PROBLEMS RELATED TO THE QUANTITATIVE ASSESSMENT OF STRETCH, ELASTIC RECOIL, AND BALLOON-ARTERY RATIO [J].
HERMANS, WRM ;
RENSING, BJ ;
STRAUSS, BH ;
SERRUYS, PW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1992, 25 (03) :174-185