Tissue resistivity estimation in the presence of positional and geometrical uncertainties

被引:10
作者
Baysal, U
Eyüboglu, BM [1 ]
机构
[1] Middle E Tech Univ, Dept Elect & Elect Engn, TR-06531 Ankara, Turkey
[2] Hacettepe Univ, Dept Elect & Elect Engn, Biomed Engn Lab, TR-06532 Ankara, Turkey
关键词
D O I
10.1088/0031-9155/45/8/322
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Geometrical uncertainties (organ boundary variation and electrode position uncertainties) are the biggest sources of error in estimating electrical resistivity of tissues from body surface measurements. In this study, in order to decrease estimation errors, the statistically constrained minimum mean squared error estimation algorithm (MiMSEE) is constrained with a priori knowledge of the geometrical uncertainties in addition to the constraints based on geometry, resistivity range, linearization and instrumentation errors. The MiMSEE calculates an optimum inverse matrix, which maps the surface measurements to the unknown resistivity distribution. The required data are obtained from four-electrode impedance measurements, similar to injected-current electrical impedance tomography (EIT). In this study, the surface measurements are simulated by using a numerical thorax model. The data are perturbed with additive instrumentation noise. Simulated surface measurements are then used to estimate the tissue resistivities by using the proposed algorithm. The results are compared with the results of conventional least squares error estimator (LSEE). Depending on the region, the MiMSEE yields an estimation error between 0.42% and 31.3% compared with 7.12% to 2010% for the LSEE. It is shown that the MiMSEE is quite robust even in the case of geometrical uncertainties.
引用
收藏
页码:2373 / 2388
页数:16
相关论文
共 22 条
[1]  
ATALAR E, 1987, THESIS MIDDLE E TU A
[2]   PRINCIPLES OF THE IMPEDANCE TECHNIQUE [J].
BAKER, LE .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1989, 8 (01) :11-15
[3]   APPLIED POTENTIAL TOMOGRAPHY [J].
BARBER, DC ;
BROWN, BH .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (09) :723-733
[4]   Use of a priori information in estimating tissue resistivities -: a simulation study [J].
Baysal, U ;
Eyüboglu, BM .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) :3589-3606
[5]  
BEST JB, 1985, BEST TAYLORS PHYSL B, P586
[6]   Imaging with electricity: Report of the European Concerted Action on Impedance Tomography [J].
Boone, K ;
Barber, D ;
Brown, B .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1997, 21 (06) :201-232
[7]   EFFECTS OF GEOMETRICAL UNCERTAINTIES ON ELECTROCARDIOGRAPHY [J].
CHOI, J ;
PILKINGTON, TC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (04) :325-334
[8]  
COMROE JH, 1963, LUNG CLIN PHYSL PULM, P323
[9]  
COTES JE, 1979, LUNG FUNCTION, P57
[10]   ESTIMATION OF TISSUE RESISTIVITIES FROM MULTIPLE-ELECTRODE IMPEDANCE MEASUREMENTS [J].
EYUBOGLU, BM ;
PILKINGTON, TC ;
WOLF, PD .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (01) :1-17