An integrated photonic sensor for in situ monitoring of hazardous organics

被引:15
作者
Luo, DH
Levy, RA
Hor, YF
Federici, JF
Pafchek, RM
机构
[1] New Jersey Inst Technol, Newark, NJ 07102 USA
[2] Agere Syst, Breinigsville, PA 18031 USA
关键词
integrated photonic sensor; Mach-Zehnder interferometry; hazardous organics;
D O I
10.1016/S0925-4005(03)00123-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study discusses the design, fabrication, and performance of an integrated photonic sensor based on the principles of single mode Mach-Zehnder interferometry. The sensor consists of a patterned waveguide structure consisting of a lower silica clad, a phosphosilicate core and an upper borophosphosilicate clad. Results indicate that such a sensor is sensitive to refractive index changes caused by the presence of small concentration of organics in water over the wavelength range of 1.25-1.65 mum. (C) 2003 Published by Elsevier Science B.V.