Stability of the Birmingham hip resurfacing arthroplasty at two years - A radiostereophotogrammetric analysis study

被引:41
作者
Itayem, R [1 ]
Arndt, A
Nistor, L
McMinn, D
Lundberg, A
机构
[1] Karolinska Univ, Karolinska Inst, Dept Orthopaed Surg, SE-14186 Stockholm, Sweden
[2] Nucl Hosp, Dept Orthopaed, SE-54185 Skovde, Sweden
[3] Birmingham Nuffield Hosp, Birmingham B15 2QQ, W Midlands, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2005年 / 87B卷 / 02期
关键词
D O I
10.1302/0301-620X.87B2.15394
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Birmingham hip resurfacing metal-on-metal arthroplasty was introduced in 1997 and has shown promising short- to mid-term results. We used radiostereophotogrammetric analysis (RSA) to study the stability of 20 resurfacing arthroplasties over a follow-up period of 24 months. Radiological examinations for RSA were performed immediately after surgery and at two, six, 12 and 24 months after operation. Precision and detection of migration thresholds (non-zero movement) were calculated. All the results corresponded well to those found in similar experimental arrangements with standard hip prostheses. Migration of the cup and vertical and mediolateral migration of the head were calculated. The values were low at two years compared with those of earlier studies of cemented femoral components in conventional total hip replacements indicating that there was no evidence of excessive early migration or loosening of the components.
引用
收藏
页码:158 / 162
页数:5
相关论文
共 16 条
[11]  
McMinn D, 1996, CLIN ORTHOP RELAT R, pS89
[12]  
MCMINN D, 2001, P AM ACAD ORTHOPAEDI
[13]  
Paltrinieri M, 1971, Chir Organi Mov, V60, P85
[14]  
Schmalzried TP, 1996, CLIN ORTHOP S, V329, P106
[15]  
TANAKA S, 1978, CLIN ORTHOP RELAT R, V134, P75
[16]  
[No title captured]