Noise reduction in PET attenuation correction using non-linear Gaussian filters

被引:19
作者
Kitamura, K
Iida, H
Shidahara, M
Miura, S
Kanno, I
机构
[1] Shimadzu Corp, Nakagyo Ku, Kyoto 6058511, Japan
[2] Natl Cardiovasc Ctr, Res Inst, Suita, Osaka 5658565, Japan
[3] Res Inst Brain & Blood Vessels, Akita 0100874, Japan
关键词
D O I
10.1109/23.856537
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In a PET study, shortening of transmission scan time is highly desired for improving patient comfort and increasing scanner throughput. It necessitates a method that reduces statistical noise in attenuation correction factors (ACFs). We have evaluated non-linear Gaussian (NLG) filtering for smoothing transmission images reconstructed with filtered back-projection instead of using iterative reconstruction and segmentation methods. The NLG filtering operation is a variation of local weighted averaging in a neighborhood around a pixel, which weights are determined according to both distance in location and difference in pixel value. Several filtering steps with different NLG parameters can effectively reduce noise without losing structural information. The NLG smoothed transmission images are then forward projected to generate ACFs. Results with phantom and patient data suggested that the NLG filtering method is useful for attenuation correction using count-limited transmission data for both brain and whole-body PET studies.
引用
收藏
页码:994 / 999
页数:6
相关论文
共 15 条
[1]  
ALENIUS S, 1998, IEEE NUCL SCI S MED
[2]  
[Anonymous], 1995, P MUSTERERKENNUNG 19
[3]   Detailed investigation of transmission and emission data smoothing protocols and their effects on emission images [J].
Chatziioannou, A ;
Dahlbom, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (01) :290-294
[4]  
Chu CK, 1998, J AM STAT ASSOC, V93, P526, DOI 10.2307/2670100
[5]   Grouped-coordinate ascent algorithms for penalized-likelihood transmission image reconstruction [J].
Fessler, JA ;
Ficaro, EP ;
Clinthorne, NH ;
Lange, K .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :166-175
[6]  
GODTLIEBSEN F, 1997, J NONPARAMETR STAT, V8, P21
[7]  
Iida H, 1996, QUANTIFICATION BRAIN, P57
[8]   Accuracy of standardized uptake value measured by simultaneous emission and transmission scanning in PET oncology [J].
Inoue, T ;
Oriuchi, N ;
Kunio, M ;
Tomiyoshi, K ;
Tomaru, Y ;
Aoyagi, K ;
Amano, S ;
Suzuki, H ;
Aoki, J ;
Sato, T ;
Endo, K .
NUCLEAR MEDICINE COMMUNICATIONS, 1999, 20 (09) :849-857
[9]   Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithm [J].
Kamphuis, C ;
Beekman, FJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (06) :1101-1105
[10]  
MEIKLE SR, 1995, J NUCL MED, V36, P1680