A high-precision system for conformal intracranial radiotherapy

被引:47
作者
Tomé, WA
Meeks, SL
Buatti, JM
Bova, FJ
Friedman, WA
Li, ZF
机构
[1] Univ Iowa, Coll Med, Div Radiat Oncol, Iowa City, IA 52242 USA
[2] Univ Wisconsin, Sch Med, Dept Human Oncol, Madison, WI USA
[3] Univ Florida, Coll Med, Dept Neurol Surg, Gainesville, FL USA
[4] Univ Florida, Coll Med, Dept Radiat Oncol, Gainesville, FL USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2000年 / 47卷 / 04期
关键词
brain neoplasms; computer-assisted radiotherapy; MRI-based radiotherapy planning; stereotaxic techniques;
D O I
10.1016/S0360-3016(00)00502-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Currently, optimally precise delivery of intracranial radiotherapy is possible with stereotactic radiosurgery and fractionated stereotactic radiotherapy. We report on an optimally precise optically guided system for three-dimensional (3D) conformal radiotherapy using multiple noncoplanar fixed fields. Methods and Materials: The optically guided system detects infrared Light emitting diodes (IRLEDs) attached to a custom bite plate linked to the patient's maxillary dentition, The IRLEDs are monitored by a commercially available stereo camera system, which is intel faced to a personal computer. An IRLED reference is established with the patient at the selected stereotactic isocenter, and the computer reports the patient's current position based on the location of the IRLEDs relative to this reference position. Using this readout from the computer, the patient may be dialed directly to the desired position in stereotactic space. The patient is localized on the first day and a reference file is established for 5 different couch positions. The patient's image data are then imported into a commercial convolution-based 3D radiotherapy planning system, The previously established isocenter and couch positions are then used as a template upon which to design a conformal 3D plan with maximum beam separation. Results: The use of the optically guided system in conjunction with noncoplanar radiotherapy treatment planning using fixed fields allows the generation of highly conformal treatment plans that exhibit a high degree of dose homogeneity and a steep dose gradient. To date, this approach has been used to treat 28 patients. Conclusion: Because IRLED technology improves the accuracy of patient localization relative to the linac isocenter and allows real-time monitoring of patient position, one can choose treatment-field margins that only account for beam penumbra and image resolution without adding margin to account for larger and poorly defined setup uncertainty. This approach enhances the normal tissue sparing, high degree of conformality, and homogeneity characteristics possible with 3D conformal radiotherapy, (C) 2000 Elsevier Science Inc.
引用
收藏
页码:1137 / 1143
页数:7
相关论文
共 29 条
[1]   OPTIMIZATION OF BEAM ORIENTATIONS IN RADIATION-THERAPY - SOME THEORETICAL CONSIDERATIONS [J].
BORTFELD, T ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (02) :291-304
[2]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[3]   Preliminary experience with frameless stereotactic radiotherapy [J].
Buatti, JM ;
Bova, FJ ;
Friedman, WA ;
Meeks, SL ;
Marcus, RB ;
Mickle, JP ;
Ellis, TL ;
Mendenhall, WM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :591-599
[4]   A comparison of three stereotactic radiotherapy techniques; arcs vs. noncoplanar fixed fields vs. intensity modulation [J].
Cardinale, RM ;
Benedict, SH ;
Wu, QW ;
Zwicker, RD ;
Gaballa, HE ;
Mohan, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02) :431-436
[5]   THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM [J].
FRIEDMAN, WA ;
BOVA, FJ .
SURGICAL NEUROLOGY, 1989, 32 (05) :334-342
[6]  
FRIEDMAN WA, 1998, LINAC RADIOSURGERY P, P1
[7]   DETERMINATION OF BEAM ORIENTATION IN RADIOTHERAPY PLANNING [J].
GOKHALE, P ;
HUSSEIN, EMA ;
KULKARNI, N .
MEDICAL PHYSICS, 1994, 21 (03) :393-400
[8]   A NONINVASIVE METHOD FOR FRACTIONATED STEREOTACTIC IRRADIATION OF BRAIN-TUMORS WITH LINEAR-ACCELERATOR [J].
HARIZ, MI ;
HENRIKSSON, R ;
LOFROTH, PO ;
LAITINEN, LV ;
SATERBORG, NE .
RADIOTHERAPY AND ONCOLOGY, 1990, 17 (01) :57-72
[9]  
Kramer B A, 1998, Radiat Oncol Investig, V6, P18, DOI 10.1002/(SICI)1520-6823(1998)6:1<18::AID-ROI3>3.3.CO
[10]  
2-J