Accuracy and precision of the TAS analyser for near-patient INR testing by non-pathology staff in the community

被引:12
作者
Cachia, PG [1 ]
McGregor, E
Adlakha, S
Davey, P
Goudie, BM
机构
[1] Dundee Teaching Hosp, NHS Trust, Ninewells Hosp & Med Sch, Dept Haematol, Dundee DD1 9SY, Scotland
[2] Univ Dundee, Med Monitoring Unit, Dundee, Scotland
[3] Westgate Hlth Ctr, Dundee, Scotland
关键词
INR; quality control; near-patient testing; Thrombolytic Assessment System analyser;
D O I
10.1136/jcp.51.1.68
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Aims-To assess the accuracy and precision of INR measurement by trained practice and district nursing staff using the Thrombolytic tic Assessment System (TAS) analyser. Methods-Seventeen nurses from four practices were trained to measure INR using the TAS analyser on citrated capillary blood samples. Quality control (QC) consisted of: daily internal QC using normal and abnormal commercial plasmas; monthly local external QC scheme using fresh citrated venous blood; and registration of all analysers in the NEQAS (national external quality assessment scheme) main users scheme. Results-Analysis of internal QC results demonstrated satisfactory interanalyser and intra-analyser precision with no evidence of analytical drift in any of the four practice analysers over an eight month period. Local and national external QC results confirmed the interanalyser precision but INR was underestimated by the TAS analysers compared with the CA 1000 using either Diagen rabbit brain thromboplastin or Innovin, and with other NEQAS users. Conclusions-The TAS analyser has many features to commend it for use by non-pathology staff to determine INR. Local internal and external QC and entry into the NEQAS main users group are possible because the TAS analyses citrated plasma or blood. The TAS analyser underestimates INR when the geometric mean normal prothrombin time (GMNPT) is determined by conventional methods. A local correction factor can be introduced by adjusting the normal PT to give INR results comparable with the local laboratory. This is particularly desirable when INRs are measured using both near-patient and laboratory analytical systems on different occasions.
引用
收藏
页码:68 / 72
页数:5
相关论文
共 10 条
[1]  
*CLIN PATH ACCR UK, 1995, ACCR HDB VERS 9 2
[2]  
COLVIN BT, 1990, J CLIN PATHOL, V43, P177
[3]  
England JM, 1995, CLIN LAB HAEMATOL, V17, P301
[4]   COST-EFFECTIVENESS OF WARFARIN AND ASPIRIN FOR PROPHYLAXIS OF STROKE IN PATIENTS WITH NONVALVULAR ATRIAL-FIBRILLATION [J].
GAGE, BF ;
CARDINALLI, AB ;
ALBERS, GW ;
OWENS, DK .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1995, 274 (23) :1839-1845
[5]   EVALUATION OF THE CIBA CORNING BIOTRACK-512 COAGULATION MONITOR FOR THE CONTROL OF ORAL ANTICOAGULATION [J].
JENNINGS, I ;
LUDDINGTON, RJ ;
BAGLIN, T .
JOURNAL OF CLINICAL PATHOLOGY, 1991, 44 (11) :950-953
[6]  
KITCHEN S, 1994, THROMB HAEMOSTASIS, V72, P426
[7]   ABC of atrital fibrillation - Antithrombotic treatment for atrial fibrillation [J].
Lip, GYH ;
Lowe, GDO .
BRITISH MEDICAL JOURNAL, 1996, 312 (7022) :45-49
[8]   EFFECT OF THROMBOPLASTIN AND COAGULOMETER INTERACTION ON THE PRECISION OF THE INTERNATIONAL NORMALIZED RATIO [J].
PI, DW ;
RABOUD, JM ;
FILBY, C ;
CARTER, CJ .
JOURNAL OF CLINICAL PATHOLOGY, 1995, 48 (01) :13-17
[9]   THE VALUE OF PLASMA CALIBRANTS IN CORRECTING COAGULOMETER EFFECTS ON INTERNATIONAL NORMALIZED RATIOS - AN INTERNATIONAL MULTICENTER STUDY [J].
POLLER, L ;
TRIPLETT, DA ;
HIRSH, J ;
CARROLL, J ;
CLARKE, K .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (03) :358-365
[10]  
WOOD E, 1994, STANDARD HAEMATOLOGY, V2, P278