The influence of surgical malalignment on the contact pressures of fixed and mobile bearing knee prostheses - a biomechanical study

被引:36
作者
Cheng, CK
Huang, CH
Liau, JJ
Huang, CH
机构
[1] Natl Yang Ming Med Univ, Inst Biomed Engn, Orthopaed Biomech Lab, Taipei 11221, Taiwan
[2] United Orthopaed Corp, Dept Res & Dev, Taipei, Taiwan
[3] Mackay Mem Hosp, Dept Orthopaed Surg, Taipei, Taiwan
关键词
mobile bearing; fixed bearing; malalignment; contact pressure; knee prosthesis;
D O I
10.1016/S0268-0033(02)00189-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To investigate the effect of surgical malalignment on contact pressures of fixed and mobile bearing knee prostheses. Design. An experimental set-up was used to measure contact pressure on the tibial component of fixed and mobile bearing knee prostheses subjected to a compression load and surgical malalignment situations were simulated. Background. Mobile bearing knee prostheses were designed to decrease tibiofemoral contact pressure by providing both high congruity and mobility. It was also assumed to accommodate surgical malalignment. However, few studies have reported the effect of malalignment of the tibiofemoral joint on contact pressure of fixed and mobile bearing knee prostheses. Methods. Surgical malalignment situations were simulated to evaluate contact characteristics of tibial component of fixed and mobile bearing knee prostheses. The simulated malalignment conditions include the medial-lateral translation (0.5 and 1 mm), anterior-posterior translation (2 and 4 mm) and internal-external rotation (1degrees, 3degrees, 5degrees and 10degrees) of the femoral component relative to the tibial component. Fuji pressure sensitive film was used to measure the contact pressure. Results. The greatest increase of maximum contact pressure in the anterior-posterior maltranslation was 7.63% and 7.62% relative to the neutral contact situation in the fixed and mobile bearing designs respectively. In the medial-lateral maltranslation, there was 23.3% in the fixed bearing design and was 22.0% in the mobile bearing design. In the internal/external malrotation, the greatest increase of maximum contact pressure in the fixed bearing design was 27.1%, which was much higher than the mobile bearing design (22.4%). Conclusions. The mobile bearing design can reduce maximum contact pressure more significantly than the fixed bearing design when malalignment conditions of the tibiofemoral joint occurs, especially in the internal/external malrotation. The mobile bearing design offers the advantage of self-adjusting over the fixed bearing design to accommodate surgical malalignment.
引用
收藏
页码:231 / 236
页数:6
相关论文
共 19 条
[1]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]   The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs [J].
Bert, JM ;
Reuben, J ;
Kelly, F ;
Gross, M ;
Elting, J .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :609-614
[3]  
BUECHEL F F, 1987, Contemporary Orthopaedics, V14, P52
[4]  
BUECHEL FF, 1989, ORTHOP CLIN N AM, V20, P147
[5]  
Cheng-Kung Cheng, 1998, Journal of Musculoskeletal Research, V2, P237, DOI 10.1142/S0218957798000238
[6]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[7]  
GOODFELLOW JW, 1986, CLIN ORTHOP RELAT R, P21
[8]   A COMPARATIVE-EVALUATION OF TIBIAL COMPONENT DESIGNS OF TOTAL KNEE PROSTHESES [J].
LEWIS, JL ;
ASKEW, MJ ;
JAYCOX, DP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (01) :129-135
[9]   The effect of malalignment on stresses in polyethylene component of total knee prostheses - a finite element analysis [J].
Liau, JJ ;
Cheng, CK ;
Huang, CH ;
Lo, WH .
CLINICAL BIOMECHANICS, 2002, 17 (02) :140-146
[10]   The influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics [J].
Liau, JJ ;
Hu, CC ;
Cheng, CK ;
Huang, CH ;
Lo, WH .
CLINICAL BIOMECHANICS, 2001, 16 (02) :160-166