摘要
Rationale and Objectives: This study aims to investigate the sensitivity of quantitative metrics derived from dynamic contrast-enhanced (DCE) magnetic resonance imaging and a pharmacokinetic (PK) model to image quality and acquisition parameters. Materials and Methods: A computer-synthesized DCE model that consisted of a large range of values of K(trans) (transfer constant of a paramagnetic contrast agent from blood to tissue), v(p) (fractional plasma volume), and k(ep) (back flux rate) was created to test the reliability of quantitative metrics derived from a standard PK model. Effects of the contrast-to-noise ratio (CNR), total acquisition time, and sampling interval on the stability and bias of the derived metrics were investigated. Results: The instability and bias of the estimated K(trans), v(p), and k(ep) values increased with sampling interval and decreased with increasing CNR. Total acquisition times had limited influence on the estimations of K(trans) and v(p) values, but increasing the total acquisition time improved the stability of the estimation of k(ep) values. However, for small k(ep) values, the stability was still poor even with a total acquisition time of 8 minutes. Also, the stability and bias of the estimated values of K(trans), v(p), and k(ep) are interrelated. Conclusions: Our synthesized DCE model represents perfectly reproduced data except for the presence of Gaussian-distributed random noise. Our analysis suggests minimum changes that may be considered potentially significant in longitudinal therapy assessment studies. Our data are complementary to experimental data from human subjects and phantoms, and provide guidance for the design of image acquisition strategies.
引用
收藏
页码:468 / 478
页数:11
相关论文
共 21 条
[1]  
Calamante F, 2000, MAGN RESON MED, V44, P466, DOI 10.1002/1522-2594(200009)44:3<466::AID-MRM18>3.0.CO
[2]  
2-M
[3]   Development of image software tools for radiation therapy assessment [J].
Cao, Y .
MEDICAL PHYSICS, 2005, 32 (06) :2136-2136
[4]   Dynamic Contrast-Enhanced Magnetic Resonance Imaging As a Biomarker for Prediction of Radiation-induced Neurocognitive Dysfunction [J].
Cao, Yue ;
Tsien, Christina I. ;
Sundgren, Pia C. ;
Nagesh, Vijaya ;
Normolle, Daniel ;
Buchtel, Henry ;
Junck, Larry ;
Lawrence, Theodore S. .
CLINICAL CANCER RESEARCH, 2009, 15 (05) :1747-1754
[5]   EARLY PREDICTION OF OUTCOME IN ADVANCED HEAD-AND-NECK CANCER BASED ON TUMOR BLOOD VOLUME ALTERATIONS DURING THERAPY: A PROSPECTIVE STUDY [J].
Cao, Yue ;
Popovtzer, Aron ;
Li, Diana ;
Chepeha, Douglas B. ;
Moyer, Jeffrey S. ;
Prince, Mark E. ;
Worden, Francis ;
Teknos, Theodoros ;
Bradford, Carol ;
Mukherji, Suresh K. ;
Eisbruch, Avraham .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (05) :1287-1290
[6]   Investigation and optimization of parameter accuracy in dynamic contrast-enhanced MRI [J].
Cheng, Hai-Ling Margaret .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (03) :736-743
[7]   Patlak plots of Gd-DTPA MRI data yield blood-brain transfer constants concordant with those of 14C-sucrose in areas of blood-brain opening [J].
Ewing, JR ;
Knight, RA ;
Nagaraja, TN ;
Yee, JS ;
Nagesh, V ;
Whitton, PA ;
Li, L ;
Fenstermacher, JD .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (02) :283-292
[8]   Reproducibility of dynamic contrast-enhanced MRI in human muscle and tumours: comparison of quantitative and semi-quantitative analysis [J].
Galbraith, SM ;
Lodge, MA ;
Taylor, NJ ;
Rustin, GJS ;
Bentzen, S ;
Stirling, JJ ;
Padhani, AR .
NMR IN BIOMEDICINE, 2002, 15 (02) :132-142
[9]   Effects of 5,6-dimethylxanthenone-4-acetic acid on human tumor microcirculation assessed by dynamic contrast-enhanced magnetic resonance imaging [J].
Galbraith, SM ;
Rustin, GJS ;
Lodge, MA ;
Taylor, NJ ;
Stirling, JJ ;
Jameson, M ;
Thompson, P ;
Hough, D ;
Gumbrell, L ;
Padhani, AR .
JOURNAL OF CLINICAL ONCOLOGY, 2002, 20 (18) :3826-3840
[10]   Temporal sampling requirements for the tracer kinetics modeling of breast disease [J].
Henderson, E ;
Rutt, BK ;
Lee, TY .
MAGNETIC RESONANCE IMAGING, 1998, 16 (09) :1057-1073