Development of a point-of-care assay system for high-sensitivity C-reactive protein in whole blood

被引:88
作者
Ahn, JS
Choi, S
Jang, SH
Chang, HJ
Kim, JH
Nahm, KB
Oh, SW
Choi, EY [1 ]
机构
[1] Hallym Univ, Dept Genet Engn, Chunchon 200702, South Korea
[2] Hallym Univ, Dept Phys, Chunchon 200702, South Korea
[3] BodiTech Inc, Cent Res Inst, Chunchon 200160, South Korea
[4] Ajou Univ, Sch Med, Dept Cardiol, Suwon 442749, South Korea
关键词
hs-CRP; fluorescence immunochromatographic assay; anti-CRP-mAb; cardiovascular diseases; laser fluorescence scanner;
D O I
10.1016/S0009-8981(03)00113-X
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: C-reactive protein (CRP) is emerging as a potential risk predictor for future cardiovascular diseases (CVD). High sensitivity assays have been developed and applied for clinical purposes. Methods: The fluorescence immunochromatographic assay was employed to detect and quantify CRP in whole blood. It consisted of a fluorescence (FL) antibody detector buffer, a test strip housed in a disposable cartridge, and a laser fluorescence scanner. Whole blood sample was mixed with detector, loaded onto a cartridge, incubated for 10 min, and the concentration of CRP was measured in a laser fluorescence scanner. The linearity, limit of detection (LOD), and performance of new assay system was tested and evaluated. The comparability of assay was examined with an automated reference method. Results: With the new assay system, a reliable correlation of coefficient (r) was obtained between the ratio value (A(T)/A(C)) and a concentration of CRP in samples. The linearity fell in the range of 0-10 mg/l of CR-P, and the analytical detection limit was 0.133 mg/l of CRP. The mean recovery of the control was 105.2% in a working range. The precision of the intra- and inter-assay in a range of 0.5-6 mg/l was CVs <6% and <8%, respectively. The new fluorescence immumochromatographic assay system correlated well with a traditional immunoturbidimetric assay for quantification of CRP concentration (r=0.955, N=90). Conclusion: The fluorescence immunochromatographic assay is fast, reliable, and a reproducible platform for point-of-care testing (POCT) of high-sensitive (hs)-CRP in whole blood. (C) 2003 Elsevier Science B.V All rights reserved.
引用
收藏
页码:51 / 59
页数:9
相关论文
共 28 条
[1]   Elevated levels of C-reactive protein at discharge in patients with unstable angina predict recurrent instability [J].
Biasucci, LM ;
Liuzzo, G ;
Grillo, RL ;
Caligiuri, G ;
Rebuzzi, AG ;
Buffon, A ;
Summaria, F ;
Ginnetti, F ;
Fadda, G ;
Maseri, A .
CIRCULATION, 1999, 99 (07) :855-860
[2]  
Borque L, 2000, CLIN CHEM, V46, P1839
[3]  
Brooks DE, 1999, CLIN CHEM, V45, P1676
[4]  
CHRISTNER RB, 1994, J BIOL CHEM, V269, P9760
[5]   Independent prognostic value of C-reactive protein and troponin I in patients with unstable angina or non-Q-wave myocardial infarction [J].
de Winter, RJ ;
Bholasingh, R ;
Lijmer, JG ;
Koster, RW ;
Gorgels, JPMC ;
Schouten, Y ;
Hoek, FJ ;
Sanders, GT .
CARDIOVASCULAR RESEARCH, 1999, 42 (01) :240-245
[6]   Production and characterization of isotype-specific monoclonal antibodies to bovine brain Rab GDI [J].
Ha, SG ;
Park, JB ;
Ko, KH ;
Choi, EY .
HYBRIDOMA, 1999, 18 (04) :371-376
[7]  
Haverkate F, 1997, LANCET, V349, P462, DOI 10.1016/S0140-6736(96)07591-5
[8]  
Hutchinson WL, 2000, CLIN CHEM, V46, P934
[9]   Simplified laser fluorescence scanner for proteomics studies and early cancer diagnosis [J].
Kim, BC ;
Jeong, JH ;
Jeong, DS ;
Choi, EY ;
Kim, JH ;
Nahm, KB .
OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS: DIAGNOSTICS AND TREATMENT, 2002, 4916 :103-108
[10]   C-reactive protein, a sensitive marker of inflammation, predicts future risk of coronary heart disease in initially healthy middle-aged men -: Results from the MONICA (Monitoring Trends and Determinants in Cardiovascular Disease) Augsburg Cohort Study, 1984 to 1992 [J].
Koenig, W ;
Sund, M ;
Fröhlich, M ;
Fischer, HG ;
Löwel, H ;
Döring, A ;
Hutchinson, WL ;
Pepys, MB .
CIRCULATION, 1999, 99 (02) :237-242