共 13 条
Micromachined separation chips with a precolumn reactor and end-column electrochemical detector
被引:103
作者:
Wang, J
[1
]
Chatrathi, MP
[1
]
Tian, BM
[1
]
机构:
[1] New Mexico State Univ, Dept Biochem & Chem, Las Cruces, NM 88003 USA
关键词:
D O I:
10.1021/ac0006371
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Glass microchips, integrating chemical derivatization reactions, electrophoretic separations, and amperometric detection, have been developed. The performance of the new integrated microfabricated devices is demonstrated for rapid on-chip measurements of amino acids utilizing precolumn reactions of amino acids with o-phthaldialdehyde/2-mercaptoethanol to generate electroactive derivatives that are separated electrophoretically and detected at the end-column electrochemical detector. The influence of the sample/reagent mixing ratio, reagent concentrations, driving voltage, detection potential, and other variables is explored. The integrated microsystem offers a rapid (6 min) simultaneous measurements of eight amino acids, down to similar to2.5 x 10(-6) M (5 fmol) level, with linearity up to the 2 x 10(-4) M level examined, and good reproducibility (RSD = 2.2-2.756), A step of the driving voltage is used for decreasing the migration time of late-eluting components and reducing the overall analysis time. The integrated microfabricated device expands the scope of on-chip electrochemical detection to nonelectro-active analytes and holds promise of being a powerful analytical tool.
引用
收藏
页码:5774 / 5778
页数:5
相关论文