Thermosensitive paramagnetic liposomes for temperature control during MR imaging-guided hyperthermia: In vitro feasibility studies

被引:74
作者
Fossheim, SL
Il'yasov, KA
Hennig, J
Bjornerud, A
机构
[1] NYCOMED AS, N-0401 Oslo, Norway
[2] Univ Freiburg, Dept Radiol, Freiburg, Germany
关键词
hyperthermia; paramagnetic liposomes; magnetic resonance; temperature monitoring; gel-to-liquid crystalline phase transition temperature; off-on switch;
D O I
10.1016/S1076-6332(00)80064-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Magnetic resonance (MR) imaging-based temperature monitoring has gained interest for use in general hyperthermia treatment of tumors. Such therapy requires an accurate control of the temperature, which should range from 41 degrees to 45 degreesC, A novel type of thermosensitive MR agent is proposed: liposome-encapsulated gadolinium chelates whose temperature response is linked to the phase-transition properties of the liposome carrier. In vitro relaxometry and MR imaging were used to evaluate the thermosensitivity of the contrast properties of liposomal gadolinium diethylenetriaminepentaacetic acid bis(methylamide) (Gd-DTPA-BMA). Materials and Methods. T1 relaxivity (r1) measurements of liposomal Gd-DTPA-BMA were undertaken at 0.47 T and at temperatures of 20 degrees -48 degreesC. MR imaging was performed at 2.0 T with a gel phantom containing inserts of liposomes. Diffusion-weighted and T1-weighted gradient-recalled echo images were acquired as the phantom was heated from 22 degrees to about 65 degreesC. Results. At ambient temperature, the r1 of liposomal Gd-DTPA-BMA was exchange limited due to slow water exchange between the liposome interior and exterior. A sharp, marked increase in r1 occurred as the temperature reached and exceeded the gel-to-liquid crystalline phase-transition temperature (T-m) of the liposomes (42 degreesC). The relaxation enhancement was mainly attributable to the marked increase in transmembrane water permeability, yielding fast exchange conditions. There was good correlation between the relaxometric and imaging results; the signal intensity on T1-weighted gradient-recalled echo images increased markedly as the temperature approached T-m. The temperature sensitivity of the diffusion-weighted technique differed from that of the liposome-based T1-weighted approach, with an apparent water diffusion coefficient increasing linearly with temperature, Conclusion. Since the transition from low to high signal intensity occurred in the temperature range of 38 degrees -42 degreesC, the investigated paramagnetic liposomes have a potential role as "off-on" switches for temperature control during hyperthermia treatment.
引用
收藏
页码:1107 / 1115
页数:9
相关论文
共 31 条
[1]   A new ytterbium chelate as contrast agent in chemical shift imaging and temperature sensitive probe for MR spectroscopy [J].
Aime, S ;
Botta, M ;
Fasano, M ;
Terreno, E ;
Kinchesh, P ;
Calabi, L ;
Paleari, L .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :648-651
[2]  
ANTICH PP, 1989, STRAHLENTHER ONKOL, V165, P734
[3]  
Chenevert Thomas L., 1995, P103
[4]   THE PROTON-RESONANCE-FREQUENCY-SHIFT METHOD COMPARED WITH MOLECULAR-DIFFUSION FOR QUANTITATIVE MEASUREMENT OF 2-DIMENSIONAL TIME-DEPENDENT TEMPERATURE DISTRIBUTION IN A PHANTOM [J].
DEPOORTER, J ;
DEWAGTER, C ;
DEDEENE, Y ;
THOMSEN, C ;
STAHLBERG, F ;
ACHTEN, E .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1994, 103 (03) :234-241
[5]  
DEPOORTER J, 1995, MAGNET RESON MED, V33, P74, DOI 10.1002/mrm.1910330111
[6]   The measurement of temperature with electron paramagnetic resonance spectroscopy [J].
Eckburg, JJ ;
Chato, JC ;
Liu, KJ ;
Grinstaff, MW ;
Swartz, HM ;
Suslick, KS ;
Auteri, FP .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (02) :193-200
[7]   Biological rationale and clinical experience with hyperthermia [J].
Engin, K .
CONTROLLED CLINICAL TRIALS, 1996, 17 (04) :316-342
[8]   Paramagnetic liposomes as MRI contrast agents: Influence of liposomal physicochemical properties on the in vitro relaxivity [J].
Fossheim, SL ;
Fahlvik, AK ;
Klaveness, J ;
Muller, RN .
MAGNETIC RESONANCE IMAGING, 1999, 17 (01) :83-89
[9]   ENCAPSULATED LIQUID-CRYSTALS AS PROBES FOR REMOTE THERMOMETRY [J].
FRANKLIN, KJ ;
BUIST, RJ ;
DENHARTOG, J ;
MCRAE, GA ;
SPENCER, DP .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (02) :253-262
[10]   Noninvasive temperature measurement in vivo using a temperature-sensitive lanthanide complex and H-1 magnetic resonance spectroscopy [J].
Frenzel, T ;
Roth, K ;
Kossler, S ;
Raduchel, B ;
Bauer, H ;
Platzek, J ;
Weinmann, HJ .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :364-369