Rapid, high sensitivity, point-of-care test for cardiac troponin based on optomagnetic biosensor

被引:88
作者
Dittmer, Wendy U. [1 ]
Evers, Toon H. [1 ]
Hardeman, Willie M. [1 ]
Huijnen, Willeke [1 ]
Kamps, Rick [1 ]
de Kievit, Peggy [1 ]
Neijzen, Jaap H. M. [1 ]
Nieuwenhuis, Jeroen H. [1 ]
Sijbers, Mara J. J. [1 ]
Dekkers, Dave W. C. [2 ]
Hefti, Marco H. [2 ]
Martens, Mike F. W. C. [2 ]
机构
[1] Philips Corp Technol, NL-5656 AE Eindhoven, Netherlands
[2] Future Diagnost BV, Wijchen, Netherlands
关键词
Magnetic particles; Electromagnet; Finger-prick blood; Dry reagents; Dynamic range; Evanescent fields; ANALYTICAL PERFORMANCE; NATIONAL ACADEMY; IMMUNOASSAY; ASSAY; GUIDELINES; CHEMISTRY; CHIP;
D O I
10.1016/j.cca.2010.03.001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: We present a prototype handheld device based on a newly developed optomagnetic technology for the sensitive detection of cardiac troponin I (cTnI) in a finger-prick blood sample with a turnaround time of 5 min. Methods: The test was completed in a compact plastic disposable with on-board dry reagents and superparamagnetic nanoparticles. In our one-step assay, all reaction processes were precisely controlled using electromagnets positioned above and below the disposable. Nanoparticle labels (500 nm) bound to the sensor surface via a sandwich immunoassay were detected using the optical technique of frustrated total internal reflection. Results: A calibration function measured in plasma demonstrates a limit of detection (mean of blank plus 3-fold the standard deviation) of 0.03 ng/mL cTnI. A linear regression analysis of the region 0.03-6.5 ng/mL yields a slope of 37 +/- 4, and a linear correlation coefficient of R(2) = 0.98. The measuring range could be extended substantially to 100 ng/mL by simultaneously imaging a second spot with a lower antibody concentration. Conclusions: The combination of magnetic particles and their fine actuation with electromagnets permits the rapid and sensitive detection of cTnI. Because of the potential high analytical performance and ease-of-use of the test, it is well suited for demanding point-of-care diagnostic applications. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:868 / 873
页数:6
相关论文
共 24 条
[1]   Validation of the 99th percentile cutoff independent of assay imprecision (CV) for cardiac troponin monitoring for ruling out myocardial infarction [J].
Apple, FS ;
Parvin, CA ;
Buechler, KF ;
Christenson, RH ;
Wu, AHB ;
Jaffe, AS .
CLINICAL CHEMISTRY, 2005, 51 (11) :2198-2200
[2]   Analytical performance of the i-STAT cardiac troponin I assay [J].
Apple, FS ;
Murakami, MM ;
Christenson, RH ;
Campbell, JL ;
Miller, CJ ;
Hock, KG ;
Scott, MG .
CLINICA CHIMICA ACTA, 2004, 345 (1-2) :123-127
[3]  
Brooks DE, 1999, CLIN CHEM, V45, P1676
[4]   Rapid integrated biosensor for multiplexed immunoassays based on actuated magnetic nanoparticles [J].
Bruls, D. M. ;
Evers, T. H. ;
Kahlman, J. A. H. ;
van Lankvelt, P. J. W. ;
Ovsyanko, M. ;
Pelssers, E. G. M. ;
Schleipen, J. J. H. B. ;
de Theije, F. K. ;
Verschuren, C. A. ;
van der Wijk, T. ;
van Zon, J. B. A. ;
Dittmer, W. U. ;
Immink, A. H. J. ;
Nieuwenhuis, J. H. ;
Prins, M. W. J. .
LAB ON A CHIP, 2009, 9 (24) :3504-3510
[5]   Cardiac point of care testing: A focused review of current National Academy of Clinical Biochemistry guidelines and measurement platforms [J].
Christenson, Robert H. ;
Azzazy, Hassan M. E. .
CLINICAL BIOCHEMISTRY, 2009, 42 (03) :150-157
[6]   Sensitive and rapid immunoassay for parathyroid hormone using magnetic particle labels and magnetic actuation [J].
Dittmer, W. U. ;
de Kievit, P. ;
Prins, M. W. J. ;
Vissers, J. L. M. ;
Mersch, M. E. C. ;
Martens, M. F. W. C. .
JOURNAL OF IMMUNOLOGICAL METHODS, 2008, 338 (1-2) :40-46
[7]   A polymer lab-on-a-chip for magnetic immunoassay with on-chip sampling and detection capabilities [J].
Do, Jaephil ;
Ahn, Chong H. .
LAB ON A CHIP, 2008, 8 (04) :542-549
[8]   Cardiac markers: a clear cause for point-of-care testing [J].
Friess, Ulrich ;
Stark, Maik .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 393 (05) :1453-1462
[10]   On-chip manipulation and detection of magnetic particles for functional biosensors [J].
Janssen, X. J. A. ;
van IJzendoorn, L. J. ;
Prins, M. W. J. .
BIOSENSORS & BIOELECTRONICS, 2008, 23 (06) :833-838