Performance evaluation of a new pulse oximeter during mild hypothermic cardiopulmonary bypass

被引:15
作者
Irita, K [1 ]
Kai, Y [1 ]
Akiyoshi, K [1 ]
Tanaka, Y [1 ]
Takahashi, S [1 ]
机构
[1] Kyushu Univ, Grad Sch Med Sci, Dept Anesthesiol & Crit Care Med, Fukuoka 8128582, Japan
关键词
D O I
10.1213/01.ANE.0000040082.50687.AB
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Newly developed pulse oximeters (POs) are designed to display accurate Spo(2), during motion and hypoperfusion. We compared the performance of a new PO, the Masimo SET Radical (M), with a conventional PO, the Nihon Kohden AY-900P (N), during hypothermic cardiopulmonary bypass. Eighteen patients were studied prospectively. PO failure was defined as failure to show no Spo(2) value or show incorrect Spo(2), values for longer than 3 min continuously. PO failure occurred in 4 and 14 patients with M and N, respectively (P = 0.0022). All 4 patients in whom PO failure developed with M were among the 14 patients with N. No Spo(2) was provided for 4% +/- 12% of the duration of aorta cross-clamping with M and 36% 39% with N (P = 0.002). Skin temperature and mean arterial blood pressure when PO failure started to occur and ended were similar between M and N. PO failure easily developed in patients with preoperative diuretic therapy or with intraoperative hyperlactatemia. in N, but not in M. M was able to display accurate Spo(2) values significantly more frequently and longer than N during mild hypothermic cardiopulmonary bypass with nonpulsatile flow, suggesting that M is more useful for monitoring Spo(2) during hypoperfusion.
引用
收藏
页码:11 / 14
页数:4
相关论文
共 19 条
[1]   The effects of motion on the performance of pulse oximeters in volunteers (revised publication) [J].
Barker, SJ ;
Shah, NK .
ANESTHESIOLOGY, 1997, 86 (01) :101-108
[2]   Continuous pulse oximetry in the general surgical ward: Nellcor N-200 versus Nellcor N-3000 [J].
Christensen, M ;
Lie, C ;
Rosenberg, J .
ANAESTHESIA, 1999, 54 (03) :253-257
[3]   Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room [J].
Dumas, C ;
Wahr, JA ;
Tremper, KK .
ANESTHESIA AND ANALGESIA, 1996, 83 (02) :269-272
[4]   PREVENTION OF INTRAOPERATIVE ANESTHESIA ACCIDENTS AND RELATED SEVERE INJURY THROUGH SAFETY MONITORING [J].
EICHHORN, JH .
ANESTHESIOLOGY, 1989, 70 (04) :572-577
[5]   Determination of systolic blood pressure via pulse oximeter in transported pediatric patients [J].
Gilmore, B ;
Hardwick, W ;
Noland, J ;
Patton, D .
PEDIATRIC EMERGENCY CARE, 1999, 15 (03) :183-186
[6]  
Goldstein MR, 1998, PEDIATRICS, V102, P818
[7]   Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults [J].
Jensen, LA ;
Onyskiw, JE ;
Prasad, NGN .
HEART & LUNG, 1998, 27 (06) :387-408
[8]   Influence of pulse oximetry and capnography on time to diagnosis of critical incidents in anesthesia: A pilot study using a full-scale patient simulator [J].
Lampotang, S ;
Gravenstein, JS ;
Euliano, TY ;
van Meurs, WL ;
Good, ML ;
Kubilis, P ;
Westhorpe, R .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1998, 14 (05) :313-321
[9]   False alarms and sensitivity of conventional pulse oximetry versus the Masimo SET™ technology in the pediatric postanesthesia care unit [J].
Malviya, S ;
Reynolds, PI ;
Voepel-Lewis, T ;
Siewert, M ;
Watson, D ;
Tait, AR ;
Tremper, K .
ANESTHESIA AND ANALGESIA, 2000, 90 (06) :1336-1340
[10]   RANDOMIZED EVALUATION OF PULSE OXIMETRY IN 20,802 PATIENTS .2. PERIOPERATIVE EVENTS AND POSTOPERATIVE COMPLICATIONS [J].
MOLLER, JT ;
JOHANNESSEN, NW ;
ESPERSEN, K ;
RAVLO, O ;
PEDERSEN, BD ;
JENSEN, PF ;
RASMUSSEN, NH ;
RASMUSSEN, LS ;
PEDERSEN, T ;
COOPER, JB ;
GRAVENSTEIN, JS ;
CHRAEMMERJORGENSEN, B ;
DJERNES, M ;
WIBERGJORGENSEN, F ;
HESLET, L ;
JOHANSEN, SH .
ANESTHESIOLOGY, 1993, 78 (03) :445-453