Time-domain combination of MR spectroscopy data acquired using phased-array coils

被引:83
作者
Brown, MA [1 ]
机构
[1] Siemens Med Solut, Siemens Training & Dev Ctr, Cary, NC 27511 USA
关键词
MR spectroscopy; phased-array coils; data processing; time domain signal processing; spectral processing;
D O I
10.1002/mrm.20244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new method for efficiently processing MRS data acquired with phased-array coils is presented. The method consists of performing phase compensation (i.e., redefining the signal phase relative to a common reference) of the signals in the time domain prior to combining the signals. The resulting spectra are equivalent to those obtained by previously published methods for phased-array spectral data processing (i.e., processing the signals individually and then combining them in the frequency domain). The method allows spectra acquired with phased-array coils to be processed as efficiently as those acquired with non-phased-array coils. Both single voxel spectroscopy (SVS) and chemical shift imaging (CSI) data sets may be processed by this method. Magn Reson Med 52:1207-1213, 2004. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:1207 / 1213
页数:7
相关论文
共 10 条
  • [1] AN NMR PHASED-ARRAY FOR HUMAN CARDIAC P-31 SPECTROSCOPY
    HARDY, CJ
    BOTTOMLEY, PA
    ROHLING, KW
    ROEMER, PB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) : 54 - 64
  • [2] NOISE CORRELATION
    JESMANOWICZ, A
    HYDE, JS
    FRONCISZ, W
    KNEELAND, JB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (01) : 36 - 47
  • [3] PROCK T, 2000, P ISMRM 8 ANN M, P562
  • [4] Pruessmann KP, 1999, MAGNET RESON MED, V42, P952, DOI 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO
  • [5] 2-S
  • [6] Fast 1H spectroscopic imaging using a multi-element head-coil array
    Schäffter, T
    Börnert, P
    Leussler, C
    Carlsen, IC
    Leibfritz, D
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (02) : 185 - 193
  • [7] Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays
    Sodickson, DK
    Manning, WJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (04) : 591 - 603
  • [8] PHASED-ARRAY DETECTORS AND AN AUTOMATED INTENSITY-CORRECTION ALGORITHM FOR HIGH-RESOLUTION MR-IMAGING OF THE HUMAN BRAIN
    WALD, LL
    CARVAJAL, L
    MOYHER, SE
    NELSON, SJ
    GRANT, PE
    BARKOVICH, AJ
    VIGNERON, DB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (03) : 433 - 439
  • [9] Wright SM, 1997, NMR BIOMED, V10, P394, DOI 10.1002/(SICI)1099-1492(199712)10:8<394::AID-NBM494>3.0.CO
  • [10] 2-0