Investigation of matrix effects for Pb isotope ratio measurements by multiple collector ICP-MS:: verification and application of optimized analytical protocols

被引:171
作者
Rehkämper, M
Mezger, K
机构
[1] Univ Munster, Inst Mineral, D-48149 Munster, Germany
[2] ETH Zurich, Inst Isotope Geol & Mineral Resources, CH-8092 Zurich, Switzerland
关键词
D O I
10.1039/b005262k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Previous studies have demonstrated the potential of multiple-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) for precise Pb isotopic measurements using admixed Tl for "external normalization" of instrumental mass discrimination. The Micromass IsoProbe is a new, single-focusing MC-ICP-MS instrument that does not employ an electrostatic sector for energy-focusing. Instead, a hexapole collision cell is used to thermalize the ion beam. This study presents the first in-depth investigation of the application of the IsoProbe to Pb isotope ratio measurements. With a Tl-based mass bias correction, multiple analyses of mixed Pb-Tl standard solutions typically display reproducibilities (+/-2 sigma) of about 50 ppm for Pb-207:Pb-206, 100 ppm for Pb-208:Pb-206 and Pb-206:Pb-204, 150 ppm for Pb-207:Pb-204 and 175 ppm for Pb-208:Pb-204. Using an empirically optimized exponential law normalization, the Pb isotope data show excellent agreement with reference values obtained by thermal ionization mass spectrometry. The cross-calibration between NIST SRM 981 and 982 demonstrates that accurate results can be obtained with this technique over a wide range of Pb isotopic compositions. The Tl-corrected Pb data obtained on the IsoProbe, however, display correlations with the Pb:Tl ratio of the analyzed solutions. The addition of large amounts of Tl to the samples is unfavorable for the accurate measurement of the low-intensity Pb-204 ion beam. Analytical artifacts may also be generated if high concentrations of concomitant elements are present in the sample solutions. Matrix effects and instrumental memory, however, can be readily overcome by the adoption of appropriate analytical protocols and it is demonstrated that sample measurements can achieve the same levels of precision and accuracy as are routinely obtained for analyses of pure standard solutions.
引用
收藏
页码:1451 / 1460
页数:10
相关论文
共 21 条
[1]   A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes [J].
Belshaw, NS ;
Freedman, PA ;
O'Nions, RK ;
Frank, M ;
Guo, Y .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1998, 181 :51-58
[2]   High precision measurement of iron isotopes by plasma source mass spectrometry [J].
Belshaw, NS ;
Zhu, XK ;
Guo, Y ;
O'Nions, RK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 197 :191-195
[3]   ABSOLUTE ISOTOPIC ABUNDANCE RATIOS OF COMMON EQUAL-ATOM AND RADIOGENIC LEAD ISOTOPIC STANDARDS [J].
CATANZARO, EJ ;
MURPHY, TJ ;
SHIELDS, WR ;
GARNER, EL .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1968, A 72 (03) :261-+
[4]   ABSOLUTE ISOTOPIC ABUNDANCE AND THE ATOMIC WEIGHT OF A REFERENCE SAMPLE OF THALLIUM [J].
DUNSTAN, LP ;
GRAMLICH, JW ;
BARNES, IL ;
PURDY, WC .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1980, 85 (01) :1-10
[5]  
Galer S. J. G., 1998, MINERAL MAG A, V62, P491, DOI DOI 10.1180/MINMAG.1998.62A.1.260
[6]  
HALLIDAY AN, 2000, MULTIPLE COLLECTOR I
[7]   Lead isotopic analyses of NIST standard reference materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effect [J].
Hirata, T .
ANALYST, 1996, 121 (10) :1407-1411
[8]   VERIFICATION OF A CORRECTION PROCEDURE FOR MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
KETTERER, ME ;
PETERS, MJ ;
TISDALE, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) :439-443
[9]  
KROGSTAD EJ, 1999, EOS T AGU, V80, pF1190
[10]   DETERMINATION OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY (ICP-MS) [J].
LONGERICH, HP ;
FRYER, BJ ;
STRONG, DF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (1-2) :39-48