We investigated the noise levels in cell-attached patch capacitance recordings with a lock-in amplifier. The capacitance noise level decreases With increasing sine wave frequency up to 20-40 kHz. With a 20-mV rms sine wave the rms noise level above 8 kHz is <50 aF. With increasing sine wave amplitudes a further reduction down to 14 aF could be achieved. Capacitance measurements with a lock-in amplifier may also be used to measure the conductance of fusion pores connecting the vesicular lumen to the extracellular space. It is estimated that at noise levels of 14 aF fusion pore conductances between 20 pS and 700 pS may be resolved in vesicles with 380-aF capacitance by using a 50-kHz sine wave. This corresponds to vesicles with a similar to 110-nm diameter. It is suggested that with low-noise techniques fusion pores may be detectable in vesicles approaching the size of large synaptic vesicles.
引用
收藏
页码:2983 / 2997
页数:15
相关论文
共 41 条
[1]   The exocytotic event in chromaffin cells revealed by patch amperometry [J].
Albillos, A ;
Dernick, G ;
Horstmann, H ;
Almers, W ;
deToledo, GA ;
Lindau, M .
NATURE, 1997, 389 (6650) :509-512
[2]   High calcium concentrations shift the mode of exocytosis to the kiss-and-run mechanism [J].
Alés, E ;
Tabares, L ;
Poyato, JM ;
Valero, V ;
Lindau, M ;
de Toledo, GA .
NATURE CELL BIOLOGY, 1999, 1 (01) :40-44
[3]   An optimized approach to membrane capacitance estimation using dual-frequency excitation [J].
Barnett, DW ;
Misler, S .
BIOPHYSICAL JOURNAL, 1997, 72 (04) :1641-1658
[4]  
Benndorf Klaus, 1995, P129
[5]   CURRENTS THROUGH THE FUSION PORE THAT FORMS DURING EXOCYTOSIS OF A SECRETORY VESICLE [J].
BRECKENRIDGE, LJ ;
ALMERS, W .
NATURE, 1987, 328 (6133) :814-817
[6]  
CURRAN MJ, 1993, J MEMBRANE BIOL, V133, P61
[7]   Influence of conductance changes on patch clamp capacitance measurements using a lock-in amplifier and limitations of the phase tracking technique [J].
Debus, K ;
Hartmann, J ;
Kilic, G ;
Lindau, M .
BIOPHYSICAL JOURNAL, 1995, 69 (06) :2808-2822
[8]   RELEASE OF SECRETORY PRODUCTS DURING TRANSIENT VESICLE FUSION [J].
DETOLEDO, GA ;
FERNANDEZCHACON, R ;
FERNANDEZ, JM .
NATURE, 1993, 363 (6429) :554-558
[9]  
DETOLEDO GA, 1987, CELL PHYSL BLOOD, P332
[10]  
DONELLY DF, 1994, BIOPHYS J, V66, P873