Drug-Elluting Stent Fracture: Incidence, Contributing Factors, and Clinical Implications

被引:62
作者
Canan, Timothy [1 ]
Lee, Michael S. [1 ]
机构
[1] Univ Calif Los Angeles, Med Ctr, Adult Cardiac Catheterizat Lab, Div Cardiol,David Geffen Sch Med, Los Angeles, CA 90095 USA
关键词
stent fracture; drug-eluting stent; restenosis; SIROLIMUS-ELUTING STENT; STRUT FRACTURE; CORONARY-STENT; INDUCED RESTENOSIS; MULTIPLE STENT; IMPLANTATION; IMPACT; ANGIOPLASTY; PREDICTORS; FREQUENCY;
D O I
10.1002/ccd.22212
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Stent fracture has been observed in noncoronary vessels, especially in the superficial femoral and popliteal arteries and with bare metal stents in saphenous vein grafts of coronary arteries. Since the introduction of drug-eluting stents, stent fractures have also been reported in small studies and case reports. We reviewed these publications to assess what is known regarding the incidence, contributing factors, and clinical implications of drug-eluting stent fracture in coronary arteries. The reported rate of drug-eluting stent fracture in coronary arteries ranges from 1 to 8%, although much of the available literature is derived from single-center studies that are heterogeneous in their study methods. A higher risk of stent fracture may be associated with the right coronary artery location, excessive tortuosity or angulation of the vessel, overlapping stents, and longer stents. The closed-cell design of the Cypher stent has been associated with increased rigidity that may increase the risk of stent fracture, although these studies did not assess the overall outcomes between the Cypher and Taxus stents in a head-to-head comparison. Stent fracture has been shown by most studies to be associated with a statistically increased incidence of focal in-stent restenosis, and some have shown an increased risk of target lesion revascularization. Other complications observed with stent fracture include stent thrombosis, coronary aneurysms, myocardial infarction, and sudden death. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:237 / 245
页数:9
相关论文
共 45 条
[1]   Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation [J].
Aoki, Jiro ;
Nakazawa, Gaku ;
Tanabe, Kengo ;
Hoye, Angela ;
Yamamoto, Hirosada ;
Nakayama, Tomohiro ;
Onuma, Yoshinobu ;
Higashikuni, Yasutomi ;
Otsuki, Syuji ;
Yagishita, Atsuhiko ;
Yachi, Sen ;
Nakajima, Hiroyoshi ;
Hara, Kazuhiro .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) :380-386
[2]  
Brilakis Emmanouil S, 2004, J Invasive Cardiol, V16, P545
[3]   Coronary-stent fracture [J].
Chowdhury, PS ;
Ramos, RG .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 347 (08) :581-581
[4]   The incidence and clinical impact of stent strut fractures developed after drug-eluting stent implantation [J].
Chung, Wook-Sung ;
Park, Chul-Soo ;
Seung, Ki-Bae ;
Kim, Pum-Joon ;
Lee, Jong-Min ;
Koo, Bon-Kwon ;
Jang, Yang-Soo ;
Yang, Joo-Young ;
Yoon, Jung-Han ;
Kim, Doo-Il ;
Yoon, Young-Won ;
Park, Jong-Seon ;
Cho, Yoon-Haeng ;
Park, Seung-Jung .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2008, 125 (03) :325-331
[5]  
DORSCH MF, 2006, J INVASIVE CARDIOL, V18, P137
[6]   Late incomplete lesion coverage following Cypher stent deployment for diffuse right coronary artery stenosis [J].
Halkin, A ;
Carlier, S ;
Leon, MB .
HEART, 2004, 90 (08) :e45
[7]  
Hamilos Michalis I, 2005, Hellenic J Cardiol, V46, P439
[8]  
JIN X, 2007, INT J CARDIOL, V118, P54
[9]   Fracture of a sirolimus-eluting stent with migration [J].
Kamura, Aoi ;
Kawasaki, Tomohiro ;
Koga, Nobuhiko ;
Inoue, Teruo ;
Node, Koichi .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2008, 130 (02) :E86-E88
[10]   Drug-eluting stent fracture occurred within 2 days after stent implantation [J].
Kang, Won Yu ;
Kim, Weon ;
Kim, Han Gyun ;
Kim, Wan .