The Otto Aufranc Award -: Highly cross-linked polyethylene in Total Hip Arthroplasty -: Randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis

被引:163
作者
Digas, G [1 ]
Kärrholm, J [1 ]
Thanner, J [1 ]
Malchau, H [1 ]
Herberts, P [1 ]
机构
[1] Univ Gothenburg, Dept Orthopaed, Inst Surg Sci, Sahlgrens Univ Hosp, S-41345 Gothenburg, Sweden
关键词
D O I
10.1097/01.blo.0000150314.70919.e3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The annual wear rate in polyethylene articulations should be 0.1 mm or less to avoid future osteolysis. Highly cross-linked polyethylene shows an 80 to 90% wear reduction in hip simulator testing, but the clinical documentation of this new polyethylene is still inadequate. We evaluated the highly crosslinked polyethylene in two prospective randomized clinical studies. Thirty-two patients (12 men, 20 women; 64 hips) with a median age of 48 years (range, 29-70 years) with bilateral primary or secondary arthrosis of the hip had hybrid total hip arthroplasty with liners made of highly crosslinked polyethylene on one side and conventional polyethylene on the other. Twenty-seven patients in this study have passed 2 years followup. A further group, comprised of 60 patients (61 hips) with a median age of 55 years (range, 35-70 years), was randomized to receive either highly cross-linked polyethylene or conventional cemented all-polyethylene of the same design. Forty-nine of these 60 patients have been observed for 3 years. In both studies all patients received Spectron stems with 28-mm Co-Cr heads. Radiostereometric examinations with the patient supine or standing were done at regular intervals. Wear was measured in the supine position from the first postoperative week, whereas standing examinations were initiated 3 months after the operation. The penetration rate almost was identical in the study and control groups at 6 months after the operation. Thereafter, the penetration rate leveled out in the two groups with highly cross-linked polyethylene. At 2 years the highly cross-linked polyethylene liner showed 62% lower proximal penetration and 31% lower total (three-dimensional) penetration when the patients were examined in supine position. The highly crosslinked all-polyethylene cemented cups showed lower proximal penetration in both positions. The better wear performance of highly cross-linked polyethylene could increase the implant longevity. Longer followup is needed to evaluate if this new material is associated with less occurrence of osteolysis.
引用
收藏
页码:6 / 16
页数:11
相关论文
共 35 条
[21]
MJOBERG B, 1990, ACTA ORTHOP SCAND, V61, P273
[22]
A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties - Recipient of the 1999 HAP Paul Award [J].
Muratoglu, OK ;
Bragdon, CR ;
O'Connor, DO ;
Jasty, M ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (02) :149-160
[23]
Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene - A new concept [J].
Muratoglu, OK ;
Bragdon, CR ;
O'Connor, D ;
Perinchief, RS ;
Estok, DM ;
Jasty, M ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :24-30
[24]
Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts [J].
Muratoglu, OK ;
Ruberti, J ;
Melotti, S ;
Spiegelberg, SH ;
Greenbaum, ES ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (07) :42-47
[25]
NASHED RS, 1995, CLIN ORTHOP RELAT R, V317, P19
[26]
Bone cement with reduced proportion of monomer in total hip arthroplasty -: Preclinical evaluation and randomized study of 47 cases with 5 years' follow-up [J].
Nivbrant, B ;
Kärrholm, J ;
Röhrl, S ;
Hassander, H ;
Wesslén, B .
ACTA ORTHOPAEDICA SCANDINAVICA, 2001, 72 (06) :572-584
[27]
Onsten I, 1996, J BONE JOINT SURG BR, V78B, P85
[28]
Wear in uncemented porous and cemented polyethylene sockets -: A randomised, radiostereometric study [J].
Önsten, I ;
Carlsson, ÅS ;
Besjakov, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02) :345-350
[29]
OONISHI H, 1992, RADIAT PHYS CHEM, V39, P495
[30]
Oonishi H, 2000, J Orthop Sci, V5, P223, DOI 10.1007/s007760050155