Validation of a small punch testing technique to characterize the mechanical behaviour of ultra-high-molecular-weight polyethylene

被引:67
作者
Kurtz, SM
Foulds, JR
Jewett, CW
Srivastav, S
Edidin, AA
机构
[1] Failure Anal Associates Inc, Philadelphia, PA 19103 USA
[2] Osteon R&D Corp, Allendale, NJ USA
关键词
ultra-high-molecular-weight polyethylene; mechanical properties; small punch test; finite; element analysis;
D O I
10.1016/S0142-9612(97)00124-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The small punch or miniaturized disc bend test has been used successfully to characterize the ductility and fracture resistance of metals and ceramics with specimens measuring 0.5 mm in thickness. This study was performed to demonstrate the feasibility of performing small punch tests on implant grade ultra-high-molecular-weight polyethylene (UHMWPE). Large-deformation finite element simulations were developed and validated to explore the hypothesis that the macroscopic constitutive behaviour of UHMWPE may be inferred from a miniature specimen testing technique which can be used to characterize the ductility and work to failure for UHMWPE. The load-displacement curve was insensitive to cyclic preconditioning of the test specimen and only mildly sensitive to the loading rate. Furthermore, the initial slope of the small punch load-displacement curve was used to determine the elastic modulus of the UHMWPE with the help of the inverse finite element method. The ultimate goal of this research is to develop the capability to perform local measurements of material tensile and static fracture properties in as-manufactured, as-sterilized and as-retrieved UHMWPE components. (C) 1998 Published by Elsevier Science Limited. All rights reserved.
引用
收藏
页码:1659 / 1663
页数:5
相关论文
共 16 条
[1]  
BOSTROM MP, 1994, CLIN ORTHOPAEDICS, V309, P20
[2]   Impact of gamma sterilization on clinical performance of polyethylene in the knee [J].
Collier, JP ;
Sperling, DK ;
Currier, JH ;
Sutula, LC ;
Saum, KA ;
Mayor, MB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (04) :377-389
[3]  
ENGELMANN B, 1991, UCRLMA105413 LAWR LI
[4]  
FOULDS JR, 1995, J TEST EVAL, V23, P3
[5]   Exponential model for the tensile true stress-strain behavior of as-irradiated and oxidatively degraded ultra high molecular weight polyethylene [J].
Kurtz, SM ;
Rimnac, CM ;
Santner, TJ ;
Bartel, DL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1996, 14 (05) :755-761
[6]   Degradation rate of ultra-high molecular weight polyethylene [J].
Kurtz, SM ;
Rimnac, CM ;
Bartel, DL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (01) :57-61
[7]  
KURTZ SM, 1994, T ORTHOP RES SOC, V19, P289
[8]  
KURTZ SM, 1997, T ASME SUM, P57
[9]  
KURTZ SM, 1997, T ASME, V36, P311
[10]  
LYKINS MD, 1995, T 21 SOC BIOM, P112