Determination of chromium(VI) and lead in water samples by on-line sorption preconcentration coupled with flame atomic absorption spectrometry using a PCTFE-beads packed column

被引:44
作者
Anthemidis, Aristidis N. [1 ]
Koussoroplis, Salome-Juliette V. [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Chem, Analyt Chem Lab, Thessaloniki 54124, Greece
关键词
chromium; lead; polychlorotrifluoroethylene; solid phase extraction; atomic absorption spectrometry; on-line; SOLID-PHASE EXTRACTION; PERFORMANCE LIQUID-CHROMATOGRAPHY; SORBENT EXTRACTION; ACTIVATED ALUMINA; AAS DETERMINATION; TRACE LEVELS; SPECIATION; SYSTEM; METALS; CR(VI);
D O I
10.1016/j.talanta.2006.08.003
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
A new time-based flow injection on-line solid phase extraction method for chromium(VI) and lead determination using flame atomic absorption spectrometry was developed. The use of hydrophobic poly-chlorotrifluoroethylene (PCTFE)-beads as absorbent in on-line preconcentration system was evaluated. Effective formation of ammonium pyrrolidine dithiocarbamate complexes and subsequently retention in PCTFE packed column, was achieved in pH range 1.0-1.6 and 1.5-3.2 for Cr(VI) and Pb(II) ions, respectively. The sorbed analyte was efficiently eluted with isobutyl-methylketone for on-line FAAS determination. The proposed packing material exhibited excellent chemical and mechanical resistance, fast kinetics for adsorption of Cr(VI) and Pb(II) permitting the use of high sample flow rates at least up to 15 mL min(-1) without loss of retention efficiency. For a preconcentration time of 90 s, the sample frequency was 30 h(-1), the enhancement factor was 94 and 220, the detection limit was 0.4 and 1.2 mu g L-1, while the precision (R.S.D.) was 1.8% (at 5 mu g L-1) and 2.1% (at 30 mu g L-1) for chromium(VI) and lead, respectively. The applicability and the accuracy of the developed method were estimated by the analysis spiked water samples and certified reference material NIST-CRM 1643d (Trace elements in water) and NIST-SRM 2109 (chromium(VI) speciation in water). (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1728 / 1733
页数:6
相关论文
共 34 条
[1]
AKITA C, 1987, ANAL SCI, V5, P473
[2]
Speciation of chromium using reversed phase-high performance liquid chromatography coupled to different spectrometric detection methods [J].
Andrle, CM ;
Jakubowski, N ;
Broekaert, JAC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (02) :189-200
[3]
[Anonymous], 1995, FLOW INJECTION ATOMI
[4]
On-line preconcentration and determination of copper, lead and chromium(VI) using unloaded polyurethane foam packed column by flame atomic absorption spectrometry in natural waters and biological samples [J].
Anthemidis, AN ;
Zachariadis, GA ;
Stratis, JA .
TALANTA, 2002, 58 (05) :831-840
[5]
Flame atomic absorption spectrometric determination of chromium(VI) by on-line preconcentration system using a PTFE packed column [J].
Anthemidis, AN ;
Zachariadis, GA ;
Kougoulis, JS ;
Stratis, JA .
TALANTA, 2002, 57 (01) :15-22
[6]
Solid phase extraction of trace elements [J].
Camel, V .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2003, 58 (07) :1177-1233
[7]
ONLINE SORBENT EXTRACTION, PRECONCENTRATION AND DETERMINATION OF LEAD BY ATOMIC-ABSORPTION SPECTROMETRY [J].
DEPENA, YP ;
GALLEGO, M ;
VALCARCEL, M .
TALANTA, 1995, 42 (02) :211-218
[8]
Flow injection solid phase extraction with Chromosorb 102:: determination of lead in soil and waters by flame atomic absorption spectrometry [J].
Elçi, L ;
Arslan, Z ;
Tyson, JF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (07) :1109-1116
[9]
Fernandes AP, 2003, ATOM SPECTROSC, V24, P179
[10]
Use of factorial design and Doehlert matrix for multivariate optimisation of an on-line preconcentration system for lead determination by flame atomic absorption spectrometry [J].
Ferreira, SLC ;
dos Santos, WNL ;
Bezerra, MA ;
Lemos, VA ;
Bosque-Sendra, JM .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 375 (05) :443-449