Use of factorial design and Doehlert matrix for multivariate optimisation of an on-line preconcentration system for lead determination by flame atomic absorption spectrometry

被引:72
作者
Ferreira, SLC
dos Santos, WNL
Bezerra, MA
Lemos, VA
Bosque-Sendra, JM
机构
[1] Univ Fed Bahia, Inst Quim, Grp Pesquisa Quim Analit, BR-40170290 Salvador, BA, Brazil
[2] Univ Estadual Sudoeste Bahia, Dept Quim & Exatas, BR-45200190 Salvador, Brazil
[3] Univ Granada, Fac Sci, Dept Analyt Chem, Granada 18071, Spain
关键词
lead determination; Doehlert matrix; on-line preconcentration; polyurethane foam; FAAS;
D O I
10.1007/s00216-002-1695-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A system for on-line preconcentration and determination of lead by flame atomic absorption spectrometry (FAAS) was proposed. It was based on the sorption of lead(II) ions on a minicolumn of polyurethane foam loaded with 2-(2-thiazolylazo)-5-dimethylaminophenol (TAM). The optimisation step was carried out using two-level full factorial and Doehlert designs for the determination of the optimum conditions for lead preconcentration. The proposed procedure allowed the determination of lead with a detection limit of 2.2 mug L-1, and a precision, calculated as relative standard deviation (RSD), of 2.4 and 6.8 for a lead concentration of 50.0 and 10.0 mug L-1, respectively. A preconcentration factor of 45 and a sampling frequency of 27 samples per hour were obtained. The recovery achieved for lead determination in the presence of several cations demonstrated that this procedure has enough selectivity for analysis of environmental samples. The validation was carried out by analysis of certified reference material. This procedure was applied to lead determination in natural food.
引用
收藏
页码:443 / 449
页数:7
相关论文
共 29 条
[1]  
[Anonymous], 1986, PRACTICAL EXPT DESIG
[2]  
[Anonymous], 1978, SPECTROCHIM ACTA, VB33, P242
[3]   Experimental design .2. Optimization [J].
Araujo, PW ;
Brereton, RG .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (02) :63-70
[4]   Experimental design .1. Screening [J].
Araujo, PW ;
Brereton, RG .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (01) :26-31
[5]  
BEVERIDGE GSG, 1971, OPTIMIZATION THEORY
[6]   OPTIMIZATION OF ANALYTICAL METHODS BY USING DOEHLERTS DESIGNS [J].
BOSQUESENDRA, JM ;
NECHAR, M ;
RODRIGUEZ, LC ;
MOLINA, MFM .
ANALYTICAL PROCEEDINGS, 1995, 32 (09) :375-377
[7]  
Box GEP., 1978, Statistics for experimenters
[8]   Multivariate optimization of solvent extraction of Cd(II), Co(II), Cr(VI), Cu(II), Ni(II), Pb(II) and Zn(II) as dibenzyldithiocarbamates and detection by AAS [J].
Camino, M ;
Bagur, MG ;
Sánchez-Viñas, M ;
Gázquez, D ;
Romero, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (06) :638-642
[9]   Polyurethane foam as a sorbent for continuous flow analysis: Preconcentration and spectrophotometric determination of zinc in biological materials [J].
de Jesus, DS ;
Casella, RJ ;
Ferreira, SLC ;
Costa, ACS ;
de Carvalho, MS ;
Santelli, RE .
ANALYTICA CHIMICA ACTA, 1998, 366 (1-3) :263-269
[10]  
DEING SN, 1987, EXPT DESIGN CHEMOMET