In-use evaluation of Perasafe® compared with Cidex® in fibreoptic bronchoscope disinfection

被引:7
作者
Hernández, A
Martró, E
Puzo, C
Matas, L
Burgués, C
Vázquez, N
Castella, J
Ausina, V
机构
[1] Univ Autonoma Barcelona, Microbiol Serv, Hosp Univ Germans Trias & Pujol, Dept Genet & Microbiol,Fac Med, Barcelona 08916, Spain
[2] Hosp Santa Creu & Sant Pau, Serv Pneumol, Barcelona, Spain
关键词
peracetic acid; alkaline glutaraldehyde; Mycobacterium tuberculosis; Mycobacterium avium-intracellulare; fibreoptic bronchoscope; disinfection;
D O I
10.1016/S0195-6701(03)00072-0
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
The mycobactericidal activity of Perasafe(R) (0.26% peracetic acid) was compared with that of Cidex(R) (2% alkaline glutaraldehyde) by an in-use test. Fibreoptic bronchoscopes were artificially contaminated with Mycobacterium tuberculosis or Mycobacterium avium-intracellulare in sputum and, after manual pre-cleaning with a neutral soap, 10 and 20 min disinfection periods were tested. Perasafe(R) was as effective as Cidex(R), thus requiring a 10 min disinfection period against M. tuberculosis and 20 min against M. avium-intracellulare. The results demonstrate that Perasafe(R) is an effective disinfectant for use in reprocessing fibreoptic bronchoscopes. (C) 2003 The Hospital Infection Society. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 29 条
  • [1] Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis - Community outbreak and nosocomial transmission via a contaminated bronchoscope
    Agerton, T
    Valway, S
    Gore, B
    Pozsik, C
    Plikaytis, B
    Woodley, C
    Onorato, I
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1997, 278 (13): : 1073 - 1077
  • [2] AYLIFFE GAJ, 1984, LANCET, V1, P161
  • [3] STANDARDIZATION OF DISINFECTANT TESTING
    AYLIFFE, GAJ
    [J]. JOURNAL OF HOSPITAL INFECTION, 1989, 13 (03) : 211 - 216
  • [4] EVALUATION OF THE STERIS-SYSTEM-1 PERACETIC-ACID ENDOSCOPE PROCESSOR
    BRADLEY, CR
    BABB, JR
    AYLIFFE, GAJ
    [J]. JOURNAL OF HOSPITAL INFECTION, 1995, 29 (02) : 143 - 151
  • [5] BRADLEY SJ, 1991, LETT APPL MICROBIOL, V13, P118
  • [6] DISINFECTANT TESTING OF CONTAMINATED ENDOSCOPES - A NEED FOR STANDARDIZATION
    CUTLER, RR
    WILSON, P
    [J]. JOURNAL OF HOSPITAL INFECTION, 1993, 25 (02) : 145 - 147
  • [7] DAWSON DJ, 1982, AM REV RESPIR DIS, V126, P1095
  • [8] FAVERO MS, 1991, INFECT CONT HOSP EP, V12, P279
  • [9] Mycobactericidal activity of selected disinfectants using a quantitative suspension test
    Griffiths, PA
    Babb, JR
    Fraise, AP
    [J]. JOURNAL OF HOSPITAL INFECTION, 1999, 41 (02) : 111 - 121
  • [10] VIRAL TRANSMISSION AND FIBEROPTIC ENDOSCOPY
    HANSON, PJV
    JEFFRIES, DJ
    COLLINS, JV
    [J]. JOURNAL OF HOSPITAL INFECTION, 1991, 18 : 136 - 140