Design studies of a high resolution PET detector using APD arrays

被引:88
作者
Shao, Y [1 ]
Silverman, RW
Farrell, R
Cirignano, L
Grazioso, R
Shah, KS
Visser, G
Clajus, M
Tümer, TO
Cherry, SR
机构
[1] Univ Calif Los Angeles, Sch Med, Dept Mol & Med Pharmacol, Crump Inst Biol Imaging, Los Angeles, CA 90024 USA
[2] Radiat Monitoring Devices Inc, Watertown, MA 02172 USA
[3] NOVA R&D Inc, Riverside, CA USA
关键词
D O I
10.1109/23.856546
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We evaluated a compact, high resolution PET detector module using avalanche photodiode (APD) arrays to replace bulky position sensitive PMTs. The newly developed APD array is a planar processed 4x4 array which has a 2x2 mm(2) pixel size with 0.4 mm gaps between pixels, about 60% quantum efficiency at 420 nm wavelength, and uniform high gain (> 1000) across all channels. A 4x4 array of 2x2x10 mm(3) LSO crystals was coupled to an APD array. Different readout electronics and signal multiplexing schemes were explored. All crystals in the detector array were clearly identified in the flood source histogram, with average peak-to-valley ratios of about 12:1 using a charge sharing resistor network. The energy resolution was measured to be similar to 14% at 511 keV in the detector array. The measured timing resolution was 2.6 ns in coincidence with a LSO/PMT detector. By optimizing the readout electronics currently being used, it is likely that detector performance can be further improved. We have also determined depth-of-interaction (DOI) by reading out two APD arrays connected to the ends of a 2x2x22 mm(3) LSO crystal. Preliminary measurements show good DOI measurement capability with DOT positioning uncertainty between 4 and 6.5 mm.
引用
收藏
页码:1051 / 1057
页数:7
相关论文
共 15 条
[1]  
CASEY ME, 1998, IEEE MIC C TOR
[2]   MicroPET: A high resolution PET scanner for imaging small animals [J].
Cherry, SR ;
Shao, Y ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Chatziioannou, A ;
Young, JW ;
Jones, WF ;
Moyers, JC ;
Newport, D ;
Boutefnouchet, A ;
Farquhar, TH ;
Andreaco, M ;
Paulus, MJ ;
Binkley, DM ;
Nutt, R ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1161-1166
[3]   Calibration of a PET detector module that measures depth of interaction [J].
Huber, JS ;
Moses, WW ;
Virador, PRG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1268-1272
[4]   Initial results from the Sherbrooke avalanche photodiode positron tomograph [J].
Lecomte, R ;
Cadorette, J ;
Rodrigue, S ;
Lapointe, D ;
Rouleau, D ;
Bentourkia, M ;
Yao, R ;
Msaki, P .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1952-1957
[5]   DESIGN AND ENGINEERING ASPECTS OF A HIGH-RESOLUTION POSITRON TOMOGRAPH FOR SMALL ANIMAL IMAGING [J].
LECOMTE, R ;
CADORETTE, J ;
RICHARD, P ;
RODRIGUE, S ;
ROULEAU, D .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1446-1452
[6]  
*LECR, 1995, HQV800 SER CHARG SEN
[7]   Design of a depth of interaction (DOI) PET detector module [J].
Miyaoka, RS ;
Lewellen, TK ;
Yu, JH ;
McDaniel, DL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1069-1073
[8]   ROOM-TEMPERATURE LSO PIN PHOTODIODE PET DETECTOR MODULE THAT MEASURES DEPTH OF INTERACTION [J].
MOSES, WW ;
DERENZO, SE ;
MELCHER, CL ;
MANENTE, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1085-1089
[9]   DESIGN STUDIES FOR A PET DETECTOR MODULE USING A PIN PHOTODIODE TO MEASURE DEPTH OF INTERACTION [J].
MOSES, WW ;
DERENZO, SE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1441-1445
[10]  
*NOVA R D INC, 1909, RENA SIGN PROC DAT S