Development of a generic microelectrode array biosensing system

被引:48
作者
Zhang, SQ [1 ]
Zhao, HJ [1 ]
John, R [1 ]
机构
[1] Griffith Univ, Fac Environm Sci, Sch Environm & Appl Sci, Gold Coast, Qld 9726, Australia
基金
澳大利亚研究理事会;
关键词
biosensor; microelectrode; array; dialysis membrane; glucose; glucose oxidase;
D O I
10.1016/S0003-2670(00)01043-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new type of electrochemical cell has been developed for use in biosensor applications. Utilising a random array of platinum microdiscs as the working electrode, it incorporates a Ag/AgCl reference electrode and platinum auxiliary electrode into the tip of a polished glassy Teflon probe. Immobilisation of the biochemical moiety onto the surface of the probe is achieved using a mixture of polyurethane (PU) and polyethylene oxide (PEO). Enclosing the tip of the probe within a dialysis membrane completes the electrochemical cell. The hydrophilic nature of the PU-PEO mixture ensures the biochemical moiety is trapped in a stable, hydrophilic environment. The result is an all in one generic electrochemical sensor that can be used in a wide range of biosensor applications in both batch or flow injection modes,The unique features of this system are demonstrated in this paper using the glucose - glucose oxidase (GOD) system. Parameters affecting sensor performance were investigated in both batch and flow injection modes. In terms of sensitivity, reproducibility, equilibration time and linear dynamic range, the performance of the microdisc based sensing system was superior to similar macroelectrode versions. Practical detection limits (0.1 mu M) were four-five orders of magnitude better for this system compared with a conventional electrode system. The advantages of using this system in flow injection mode, particularly in terms of increased linear range, were also addressed. The overall linear range of our system (0.1 mu M-60 mM) is three-four orders of magnitude greater than the conventional electrode system. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:175 / 187
页数:13
相关论文
共 40 条
[11]   MICROELECTRODE ARRAYS AND APPLICATION TO BIOSENSING DEVICES [J].
HINTSCHE, R ;
PAESCHKE, M ;
WOLLENBERGER, U ;
SCHNAKENBERG, U ;
WAGNER, B ;
LISEC, T .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (9-10) :697-705
[12]   DEVELOPMENT OF CHOLINE AND ACETYLCHOLINE PT MICROELECTRODES [J].
HUANG, ZX ;
VILLARTASNOW, R ;
LUBRANO, GJ ;
GUILBAULT, GG .
ANALYTICAL BIOCHEMISTRY, 1993, 215 (01) :31-37
[13]  
John R, 1998, ELECTROANAL, V10, P1125
[14]   DOPING DEDOPING OF POLYPYRROLE - A STUDY USING CURRENT-MEASURING AND RESISTANCE-MEASURING TECHNIQUES [J].
JOHN, R ;
WALLACE, GG .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1993, 354 (1-2) :145-160
[15]   DISPERSED MERCURY MICROELECTRODES USING NONCONDUCTING POLYMER-COATINGS [J].
JOHN, R ;
WALLACE, GG .
ANALYTICA CHIMICA ACTA, 1990, 235 (02) :451-455
[16]  
Kirk J K, 1998, J Am Pharm Assoc (Wash), V38, P210
[17]   Microfabricated biosensors and microanalytical systems for blood analysis [J].
Lauks, IR .
ACCOUNTS OF CHEMICAL RESEARCH, 1998, 31 (05) :317-324
[18]   Polymers and enzyme biosensors [J].
Liu, YC ;
Yu, TY .
JOURNAL OF MACROMOLECULAR SCIENCE-REVIEWS IN MACROMOLECULAR CHEMISTRY AND PHYSICS, 1997, C37 (03) :459-500
[19]   Theoretical treatment of the current vs. time response of microelectrode arrays to changes of potential, concentration, or flow [J].
Morf, WE .
ANALYTICA CHIMICA ACTA, 1997, 341 (2-3) :121-127
[20]   Theoretical treatment of the amperometric current response of multiple microelectrode arrays [J].
Morf, WE .
ANALYTICA CHIMICA ACTA, 1996, 330 (2-3) :139-149