New direct optical biosensors for multi-analyte detection

被引:39
作者
Nikitin, PI [1 ]
Valeiko, MV [1 ]
Gorshkov, BG [1 ]
机构
[1] Russian Acad Sci, Inst Gen Phys, Moscow 119991, Russia
关键词
optics; interference; phase measurements; biosensors; array;
D O I
10.1016/S0925-4005(03)00020-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new label-free optical biosensing method has been developed, which can be called Spectral Correlation Method (SCM). The method is based on measurements of a correlation signal of two interferometers in an original optical scheme. The first interferometer is a scanned Fabry-Perot interferometer, in which the distance between two mirrors is periodically changed by a peizoelectric driver. The second interferometer is formed by a biochip, which is a simple glass plate with thickness of several tens or hundreds microns with different recognition spots or wells, or flow channels. In our experiments, a microscope cover glass was successfully used as a biochip to detect reactions of binding or detachment of different biological agents in real time. Two- and 96-channel SCM devices have been designed to register bio- and chemical-components in real time. The sensogram drifts for these devices expressed in terms of layer thickness were 20 and 40 pm/h, respectively. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 13 条
[1]   THE RESONANT MIRROR - A NOVEL OPTICAL BIOSENSOR FOR DIRECT SENSING OF BIOMOLECULAR INTERACTIONS .1. PRINCIPLE OF OPERATION AND ASSOCIATED INSTRUMENTATION [J].
CUSH, R ;
CRONIN, JM ;
STEWART, WJ ;
MAULE, CH ;
MOLLOY, J ;
GODDARD, NJ .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (7-8) :347-353
[2]   CHEMICAL AND BIOCHEMICAL SENSORS BASED ON INTERFEROMETRY AT THIN (MULTI-)LAYERS [J].
GAUGLITZ, G ;
BRECHT, A ;
KRAUS, G ;
NAHM, W .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 11 (1-3) :21-27
[3]   Dark-field surface plasmon resonance microscopy [J].
Grigorenko, AN ;
Beloglazov, AA ;
Nikitin, PI ;
Kuhne, C ;
Steiner, G ;
Salzer, R .
OPTICS COMMUNICATIONS, 2000, 174 (1-4) :151-155
[4]   Remote opto-chemical sensing with extreme sensitivity: design, fabrication and performance of a pigtailed integrated optical phase-modulated Mach-Zehnder interferometer system [J].
Heideman, RG ;
Lambeck, PV .
SENSORS AND ACTUATORS B-CHEMICAL, 1999, 61 (1-3) :100-127
[5]   Phase-polarisation contrast for surface plasmon resonance biosensors [J].
Kabashin, AV ;
Kochergin, VE ;
Beloglazov, AA ;
Nikitin, PI .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (12) :1263-1269
[6]  
Liedberg B., 1995, BIOSENS BIOELECTRON, V10, pI
[7]  
LUKOSZ W, 1991, SENSOR ACTUAT A-PHYS, V25, P185
[8]  
NIKITIN P, 1997, P TRANSDUCERS 97 CHI, P495
[9]   Surface plasmon resonance interferometry for biological and chemical sensing [J].
Nikitin, PI ;
Beloglazov, AA ;
Kochergin, VE ;
Valeiko, MV ;
Ksenevich, TI .
SENSORS AND ACTUATORS B-CHEMICAL, 1999, 54 (1-2) :43-50
[10]   Spectral-phase interference method for detecting biochemical reactions on a surface [J].
Nikitin, PI ;
Gorshkov, BG ;
Valeiko, MV ;
Rogov, SI .
QUANTUM ELECTRONICS, 2000, 30 (12) :1099-1104