Objective: Computed tomographic angiography (CTA) is currently the most commonly used technique for postoperative surveillance to detect endoleaks after endovascular stent graft repair of abdominal aortic aneurysms. We have evaluated the efficacy of duplex ultrasound scan with the addition of an ultrasound scan contrast agent in documenting endoleaks and compared these results with CTA. Methods: Conventional duplex ultrasound scan with color Doppler imaging (CDI) was first done as part of routine postoperative surveillance. After the CDI study, a 1-mL bolus of ultrasound scan contrast was given via an antecubital vein, followed by a 5-mL flush with normal saline solution. The duplex ultrasound scan evaluation of the stent graft was repeated once the contrast agent was circulating throughout the blood pool, with tissue harmonic imaging to optimize visualization of the contrast agent. Status of the stent graft, the presence or absence of any endoleak, and whether these endoleaks were graft related (group 1) or arterial branch related (group 11) were recorded. Findings were compared with CTA studies done within a 2-week period of the ultrasound scan examination. Results. Twenty patients were evaluated, 18 with modular stent grafts and two with unibody bifurcated stent grafts. Patients had a mean age of 74.5 +/- 7.6 years; 19 were male, with only one female. All stent grafts remained widely patent with normal aortoiliac flow hemodynamics. Duplex ultrasound scan with contrast identified all eight of the endoleaks seen with CTA and was able to determine whether they were group I or group 11 leaks. In two patients, ultrasound scan with contrast detected small endoleaks at the proximal graft attachment site, with extravasation of contrast into the aneurysm sac during systole. These endoleaks were not seen with CTA but were confirmed with conventional angiography at the time of endovascular closure. Standard duplex ultrasound scan with CDI failed to identify four of the 10 endoleaks in patients with technically difficult conditions. Conclusion: Duplex ultrasound scan, when used with an intravenously administered ultrasound scan contrast agent in the noninvasive follow-up of patients with aortic stent grafts, appears to provide good sensitivity to the presence and type of endoleaks, even in patients with technically difficult conditions not amenable to conventional duplex ultrasound scan with CDI.
引用
收藏
页码:381 / 385
页数:5
相关论文
共 22 条
[1]   Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, RP ;
Cope, C ;
Golden, MA ;
Velazquez, OC ;
Neschis, DG ;
Mitchell, ME ;
Barker, CF ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :32-40
[2]   Predicting the acoustic response of a microbubble population for contrast imaging in medical ultrasound [J].
Chin, CT ;
Burns, PN .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (08) :1293-1300
[3]   Endoleak after endovascular repair of abdominal aortic aneurysm [J].
Chuter, TAM ;
Faruqi, RM ;
Sawhney, R ;
Reilly, LM ;
Kerlan, RB ;
Canto, CJ ;
Lukaszewicz, GC ;
LaBerge, JM ;
Wilson, MW ;
Gordon, RL ;
Wall, SD ;
Rapp, J ;
Messina, LM .
JOURNAL OF VASCULAR SURGERY, 2001, 34 (01) :98-105
[4]  
Clark LN, 2000, AM J CARDIOL, V86, p14G
[5]   Follow-up evaluation of endoluminally treated abdominal aortic aneurysms with duplex ultrasonography: Validation with computed tomography [J].
d'Audiffret, A ;
Desgranges, P ;
Kobeiter, DH ;
Becquemin, JP .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :42-49
[6]   Detection procedures of ultrasound contrast agents [J].
de Jong, N ;
Frinking, PJA ;
Bouakaz, A ;
Ten Cate, FJ .
ULTRASONICS, 2000, 38 (1-8) :87-92
[7]  
Dowling R J, 1999, Australas Radiol, V43, P206
[8]   Ultrasound contrast imaging: Current and new potential methods [J].
Frinking, PJA ;
Bouakaz, A ;
Kirkhorn, J ;
Ten Cate, FJ ;
de Jong, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (06) :965-975
[9]   Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experience [J].
Harris, PL ;
Vallabhaneni, SR ;
Desgranges, P ;
Bacquemin, JP ;
van Marrewijk, C ;
Laheij, RJF .
JOURNAL OF VASCULAR SURGERY, 2000, 32 (04) :739-749
[10]  
JOHNSON BL, 1998, J VASC TECHNOL, V22, P97