A remotely requestable, passive, short-range sensor network for measuring small voltages is presented. The sensor system is able to simultaneously monitor six small voltages in millivolt-range, and it can be used for Holter-electrocardiogram (ECG) and other biopotential monitoring, or in industrial applications. The sensors are based on a surface acoustic wave (SAW) delay line with voltage-dependent, impedance loading on a reflector interdigital transducer (IDT). The load circuit impedance is varied by the capacitance of the voltage-controlled varactor. High resolution is achieved by developing a MOS-capacitor with a thin oxide, low flat-band voltage, and zero-voltage capacitance in the space-charge region, as well as a high-Q-microcoil by thick metal electroplating. Simultaneous monitoring of multiple potentials is realized by time-division-multiplexing of different sensor signals.
引用
收藏
页码:1464 / 1469
页数:6
相关论文
共 13 条
[1]   Comparative studies on polymer coated SAW and STW resonators for chemical gas sensor applications [J].
Avramov, ID ;
Rapp, M ;
Voigt, A ;
Stahl, U ;
Dirschka, M .
PROCEEDINGS OF THE 2000 IEEE/EIA INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM & EXHIBITION, 2000, :58-65
[2]   The mobile patient: Wireless distributed sensor networks for patient monitoring and care [J].
Bauer, P ;
Sichitiu, M ;
Istepanian, R ;
Premaratne, K .
2000 IEEE EMBS INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY APPLICATIONS IN BIOMEDICINE, PROCEEDINGS, 2000, :17-21
[3]   Ergebnisse der EKG-Interpretation mittels SignalmustererkennungResults of ECG interpretation by means of signal waveform recognition [J].
R. Bousseljot ;
D. Kreiseler .
Herzschrittmachertherapie + Elektrophysiologie, 2000, 11 (4) :197-206
[4]   Computationally efficient ECG compression scheme using a non-linear quantizer [J].
Cassen, M ;
English, MJ .
COMPUTERS IN CARDIOLOGY 1997, VOL 24, 1997, 24 :283-286
[5]   Implementation of a WAP-Based telemedicine system for patient monitoring [J].
Hung, K ;
Zhang, YT .
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2003, 7 (02) :101-107
[6]   In-home wireless monitoring of physiological data for heart failure patients [J].
Mendoza, GG ;
Tran, BQ .
SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, :1849-1850
[7]  
MORIIUZUMI T, 1994, P IEEE ULTR S NOV, P499
[8]  
MULLER J, 2002, Patent No. 10361991
[9]   A review of wireless SAW sensors [J].
Pohl, A .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (02) :317-332
[10]  
Schimetta G, 2000, IEEE T MICROW THEORY, V48, P2730, DOI 10.1109/22.899037