Least detectable concentration and dynamic range of three immunoassay systems using the same antibody

被引:25
作者
Glass, Thomas R. [1 ]
Ohmura, Naoya [1 ]
Saiki, Hiroshi [1 ]
机构
[1] Cent Res Inst Elect Power Ind, Dept Biosci, Chiba 2701194, Japan
关键词
D O I
10.1021/ac061288z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The least detectable concentration (LDC) and dynamic range (DR) of three immunoassay systems are compared using four distinct antibodies (all specific for estradiol but with different affinities) on each system. The systems evaluated include the industry standard, ELISA, and two biosensors, surface plasmon resonance and kinetic exclusion. In all cases, the measurements of inhibition curves (response vs estradiol concentration) were contracted to outside experts (the biosensor manufacturers themselves in the case of the biosensors), each of whom was supplied with the same blind samples. Each biosensor manufacturer also reported an estimate of the equilibrium dissociation constant (K-d) for each of the antibodies. The LDC and DR observed for the kinetic exclusion biosensor are consistent with an interpretation of K-d limited detection while that from the other biosensor and ELISA show limits of detection somewhat above those expected for K-d limited performance. The determined LDC and DR of each biosensor is self-consistent in the sense that none of the inhibition data contradicts theoretical limits associated with the K-d as measured on that system; however, some contradictions are apparent across platforms. The use of multiple antibodies of differing K-d improves confidence that the observed differences in performance are associated with the immunoassay system rather than the particular analyte.
引用
收藏
页码:1954 / 1960
页数:7
相关论文
共 8 条
[1]  
Ekins R, 1997, CLIN CHEM, V43, P1824
[2]   BASIC PRINCIPLES AND THEORY [J].
EKINS, RP .
BRITISH MEDICAL BULLETIN, 1974, 30 (01) :3-11
[3]  
Ekins RP, 1998, CLIN CHEM, V44, P2015
[4]  
GOSLING JP, 1990, CLIN CHEM, V36, P1408
[5]   Precision, limit of detection and range of quantitation in competitive ELISA [J].
Hayashi, Y ;
Matsuda, R ;
Maitani, T ;
Imai, K ;
Nishimura, W ;
Ito, K ;
Maeda, M .
ANALYTICAL CHEMISTRY, 2004, 76 (05) :1295-1301
[6]   THE DEPENDENCE OF THE DETECTION LIMIT OF REAGENT-LIMITED IMMUNOASSAY ON ANTIBODY-AFFINITY [J].
OCONNOR, T ;
KANE, MM ;
GOSLING, JP .
BIOCHEMICAL SOCIETY TRANSACTIONS, 1995, 23 (02) :S393-S393
[7]   An immunoassay for small analytes with theoretical detection limits [J].
Ohmura, N ;
Lackie, SJ ;
Saiki, H .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3392-3399
[8]  
Pardue HL, 1997, CLIN CHEM, V43, P1831