HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF 4,4'-METHYLENEDIANILINE IN HUMAN URINE

被引:9
作者
PETERSON, JC
ESTIVA, EC
LYTTLE, DS
HARRIS, RM
机构
[1] Pacific Toxicology Laboratories, Los Angeles, CA 90025
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1991年 / 564卷 / 01期
关键词
D O I
10.1016/0378-4347(91)80082-N
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is described for the determination of urinary 4,4'-methylenedianiline (MDA) by high-performance liquid chromatography (HPLC). MDA was extracted from hydrolyzed urine using C18 solid-phase extraction columns. The extract was analyzed by reversed-phase HPLC with electrochemical detection at a cell potential of 0.8 V. The method was very sensitive (detection limit 2.5-mu-g/l) and quantitation using 4,4'-ethylenedianiline as an internal standard correlated well with results by gas chromatography - mass spectrometry. Run-to-run precision (n = 25) averaged 8.9%. In analysis of more than 160 potentially exposed workers, MDA was detected in less than 20% of the urines and concentrations ranged up to 210-mu-g MDA per g of creatinine.