SINGLE-PHOTON AVALANCHE-DIODE FOR SINGLE-MOLECULE DETECTION

被引:123
作者
LI, LQ
DAVIS, LM
机构
关键词
D O I
10.1063/1.1144463
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A commercially available single photon avalanche photodiode in a passively quenched circuit is used with time-correlated single photon counting modules to achieve subnanosecond time response together with high quantum efficiency and low dark noise. These characteristics are required for experiments in single molecule detection and spectroscopy in which time-gated detection schemes are used. By tightly focusing the input light onto the active area, a quantum efficiency of over 50% and a single photon timing jitter of 168 ps full width at half-maximum are achieved. In addition, the full width at one-hundredth maximum, which is of greater importance for time gating, is 790 ps, comparable to that from a microchannel plate photomultiplier. Measurements of the detector dead time, and the quantum efficiency, dark counts, time response, and pulse height distribution for different operating conditions are also reported.
引用
收藏
页码:1524 / 1529
页数:6
相关论文
共 20 条
[1]   CHARACTERIZATION OF SILICON AVALANCHE PHOTODIODES FOR PHOTON-CORRELATION MEASUREMENTS .2. ACTIVE QUENCHING [J].
BROWN, RGW ;
JONES, R ;
RARITY, JG ;
RIDLEY, KD .
APPLIED OPTICS, 1987, 26 (12) :2383-2389
[2]   CHARACTERIZATION OF SILICON AVALANCHE PHOTODIODES FOR PHOTON-CORRELATION MEASUREMENTS .1. PASSIVE QUENCHING [J].
BROWN, RGW ;
RIDLEY, KD ;
RARITY, JG .
APPLIED OPTICS, 1986, 25 (22) :4122-4126
[3]   ALL-SOLID-STATE MICROSCOPE-BASED SYSTEM FOR PICOSECOND TIME-RESOLVED PHOTOLUMINESCENCE MEASUREMENTS ON II-VI SEMICONDUCTORS [J].
BULLER, GS ;
MASSA, JS ;
WALKER, AC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (05) :2994-2998
[4]   20-PS TIMING RESOLUTION WITH SINGLE-PHOTON AVALANCHE-DIODES [J].
COVA, S ;
LACAITA, A ;
GHIONI, M ;
RIPAMONTI, G ;
LOUIS, TA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (06) :1104-1110
[5]   TOWARDS PICOSECOND RESOLUTION WITH SINGLE-PHOTON AVALANCHE-DIODES [J].
COVA, S ;
LONGONI, A ;
ANDREONI, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (03) :408-412
[6]   A SEMICONDUCTOR DETECTOR FOR MEASURING ULTRAWEAK FLUORESCENCE DECAYS WITH 70 PS FWHM RESOLUTION [J].
COVA, S ;
LONGONI, A ;
ANDREONI, A ;
CUBEDDU, R .
IEEE JOURNAL OF QUANTUM ELECTRONICS, 1983, 19 (04) :630-634
[7]   IMPROVING THE PERFORMANCE OF COMMERCIALLY AVAILABLE GEIGER-MODE AVALANCHE PHOTODIODES [J].
GHIONI, M ;
RIPAMONTI, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (01) :163-167
[8]   PHOTON-COUNTING WITH PHOTO-DIODES [J].
INGERSON, TE ;
KEARNEY, RJ ;
COULTER, RL .
APPLIED OPTICS, 1983, 22 (13) :2013-2018
[9]   ULTRAFAST MICROCHANNEL PLATE PHOTOMULTIPLIERS [J].
KUME, H ;
KOYAMA, K ;
NAKATSUGAWA, K ;
SUZUKI, S ;
FATLOWITZ, D .
APPLIED OPTICS, 1988, 27 (06) :1170-1178
[10]   DOUBLE EPITAXY IMPROVES SINGLE-PHOTON AVALANCHE-DIODE PERFORMANCE [J].
LACAITA, A ;
GHIONI, M ;
COVA, S .
ELECTRONICS LETTERS, 1989, 25 (13) :841-843