TOWARDS UNDERSTANDING THE LIMITS OF DISTORTION-PRODUCT OTOACOUSTIC EMISSION MEASUREMENTS

被引:43
作者
GORGA, MP
NEELY, ST
BERGMAN, BM
BEAUCHAINE, KL
KAMINSKI, JR
LIU, Z
机构
[1] Boys Town National Research Hospital, Omaha, Nebraska 68131
关键词
D O I
10.1121/1.410227
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Distortion product otoacoustic emission (DPOAE) data were obtained with a custom-designed system from 20 subjects with normal hearing. Cavity measurements, using this system and an Etymotic ER-10B low-noise microphone system, resulted in estimates of recording system distortion of -20 dB SPL for f(2) frequencies ranging from 500 to 8000 Hz, and primary levels ranging from 20 to 75 dB SPL (L(2) = L(1) - 10 dB). Using this system it was possible to automatically adjust averaging time in order to obtain the same residual noise levels across frequencies. In all subjects with normal hearing, DPOAEs were measurable over a wide range of primary levels for octave f(2) frequencies from 1000 to 8000 Hz, but not at 500 Hz. At 500 Hz, only half of the normal-hearing subjects produced DPOAEs that were above the noise floor. When they did, DPOAE amplitude was less than that observed at higher f(2) frequencies. While the cause for response absence in some normal ears may have been due to residual noise, the reduced amplitude suggests that the other factors influence the measurement of DPOAEs at low frequencies. This result may be due to reduced cochlear production of DPOAEs at lower frequencies or reduced transmission through the middle ear.
引用
收藏
页码:1494 / 1500
页数:7
相关论文
共 13 条
[1]  
ALLEN JB, 1990, USER MANUAL CUBDIS D
[2]  
ALLEN JB, 1994, COMMUNICATION
[3]   THE BEHAVIOR OF THE ACOUSTIC DISTORTION PRODUCT, 2F1-F2, FROM THE HUMAN EAR AND ITS RELATION TO AUDITORY-SENSITIVITY [J].
GASKILL, SA ;
BROWN, AM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (02) :821-839
[4]   A COMPARISON OF TRANSIENT-EVOKED AND DISTORTION-PRODUCT OTOACOUSTIC EMISSIONS IN NORMAL-HEARING AND HEARING-IMPAIRED SUBJECTS [J].
GORGA, MP ;
NEELY, ST ;
BERGMAN, BM ;
BEAUCHAINE, KL ;
KAMINSKI, JR ;
PETERS, J ;
SCHULTE, L ;
JESTEADT, W .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 94 (05) :2639-2648
[5]   OTOACOUSTIC EMISSIONS FROM NORMAL-HEARING AND HEARING-IMPAIRED SUBJECTS - DISTORTION PRODUCT RESPONSES [J].
GORGA, MP ;
NEELY, ST ;
BERGMAN, B ;
BEAUCHAINE, KL ;
KAMINSKI, JR ;
PETERS, J ;
JESTEADT, W .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 93 (04) :2050-2060
[6]   DISTORTION PRODUCT EMISSIONS IN HUMANS .1. BASIC PROPERTIES IN NORMALLY HEARING SUBJECTS [J].
LONSBURYMARTIN, BL ;
HARRIS, FP ;
HAWKINS, MD ;
STAGNER, BB ;
MARTIN, GK .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1990, 99 (05) :3-14
[7]   DISTORTION PRODUCT EMISSIONS IN HUMANS .3. INFLUENCE OF SENSORINEURAL HEARING-LOSS [J].
MARTIN, GK ;
OHLMS, LA ;
HARRIS, FP ;
FRANKLIN, DJ ;
LONSBURYMARTIN, BL .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1990, 99 (05) :30-42
[8]   DISTORTION-PRODUCT EMISSIONS AND AUDITORY-SENSITIVITY IN HUMAN EARS WITH NORMAL HEARING AND COCHLEAR HEARING-LOSS [J].
NELSON, DA ;
KIMBERLEY, BP .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1992, 35 (05) :1142-1159
[9]   ANALYSIS OF TRANSIENT-EVOKED OTOACOUSTIC EMISSIONS IN NORMAL-HEARING AND HEARING-IMPAIRED EARS [J].
PRIEVE, BA ;
GORGA, MP ;
SCHMIDT, A ;
NEELY, S ;
PETERS, J ;
SCHULTES, L ;
JESTEADT, W .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 93 (06) :3308-3319
[10]   THE INFLUENCE OF NOISE ON THE MEASURED AMPLITUDES OF DISTORTION-PRODUCT OTOACOUSTIC EMISSIONS [J].
WHITEHEAD, ML ;
LONSBURYMARTIN, BL ;
MARTIN, GK .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1993, 36 (05) :1097-1102