RADIATION-RESISTANCE OF MICROORGANISMS ON UNSTERILIZED INFUSION SETS

被引:2
作者
CHRISTENSEN, EA
KRISTENSEN, H
HOBORN, J
MILLER, A
机构
[1] RISO NATL LAB,DEPT ACCELERATOR,DK-4000 ROSKILDE,DENMARK
[2] MOLNLYCKE HLTH CARE AB,GOTHENBURG,SWEDEN
关键词
RADIATION RESISTANCE; STERILIZATION; MEDICAL DEVICE; DEINOCOCCUS-RADIODURANS;
D O I
10.1111/j.1699-0463.1991.tb01236.x
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Three different methods were used for detecting and isolating microorganisms with high radiation resistance from the microbial contamination on infusion sets prior to sterilization. By all three methods, microorganisms with a radiation resistance high enough to be a critical factor in a sterilization process (D-6 value greater-than-or-equal-to 30 kGy) were found with a frequency of approximately two colony forming units (cfu) per 100 product items, even though the product items in two of the series of analyses were irradiated with doses of 3-6 kGy. The frequency of occurrence of isolates with D-6 values greater-than-or-equal-to 30 kGy was 0.45 per 1000 cfu of the total aerobic count. Eight different isolates of microorganisms had D-6 values greater-than-or-equal-to 40 kGy when irradiated in dried laboratory preparation. All but one of these were classified according to morphologic criteria as Deinococcus, and all but one had nonlinear dose-response relationships in semilogarithmic presentation.
引用
收藏
页码:620 / 626
页数:7
相关论文
共 13 条
[11]  
1984, PROCESS CONTROL GUID
[12]  
1967, RAD STERILIZATION ME, P423
[13]  
1987, RADIAT PHYS CHEM, V29, P87