CONTRAST ANALYSIS OF CRYO-IMAGES OF N-PARAFFIN RECORDED AT 400 KV OUT TO 2.1 A RESOLUTION

被引:31
作者
BRINK, J
CHIU, W
机构
[1] Verna and Marrs McLean Department of Biochemistry, Baylor College of Medicine, Houston, Texas, 77030, One Baylor Plaza
关键词
400-KEV ELECTRONS; HIGH-RESOLUTION ELECTRON CRYOMICROSCOPY; IMAGE CONTRAST; INELASTIC SCATTERING; MICROSCOPE ENVELOPE FUNCTION; MODULATION TRANSFER FUNCTION; BEAM-INDUCED SPECIMEN MOVEMENT;
D O I
10.1111/j.1365-2818.1991.tb03090.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
n-Paraffin was used as a test specimen for evaluating the relative merits of 400-kV versus 100-kV electron microscopy in recording data for electron crystallographic analysis of beam-sensitive materials. The parameter used for comparison, the relative contrast R, is the ratio of amplitudes from the computed Fourier transform of images and amplitudes from an electron diffraction pattern from the same crystal. R will thus be a measure of the contrast from an experimental image relative to that of a perfect image. Electron diffraction patterns and bright-field images were recorded at 400 kV at a specimen temperature of - 167-degrees-C. Using the flood-beam imaging technique the best R-value is 0.08 for all reflections in the resolution zone from 4 to 3 angstrom. This value is equivalent to that found at 100 kV. In the resolution zone from 3 to 2 angstrom we have found R = 0.02. Using the spot-scan imaging technique, on the other hand, R was measured to be 0.42 for the reflections between 4- and 3-angstrom resolution. This amount of relative contrast is 1.7 times that observed at 100 kV. Reflections at 3-2 angstrom displayed an R-value of 0.05. Besides obtaining higher R-values when applying the spot-scan imaging technique at 400 kV, we observe a higher yield of images with isotropic diffraction and/or higher resolution reflections. Various contrast-attenuating factors, including the modulation transfer function of the photographic film and the cryoholder, envelope functions for spatial and temporal coherence and lens and hightension instabilities, the contrast transfer function and lastly the radiation damage effects, have been considered in interpreting the observed image contrast. Overall, use of 400 kV in combination with spot-scan does offer important improvements in contrast levels, which can be very useful in determining the three-dimensional structure from protein crystals.
引用
收藏
页码:279 / 295
页数:17
相关论文
共 38 条